Comparison of Vacuum Pressures and Forces Generated by Different Catheters and Pumps for Aspiration Thrombectomy in Acute Ischemic Stroke
Open Access
- 18 May 2017
- journal article
- research article
- Published by S. Karger AG in Interventional Neurology
- Vol. 6 (3-4) , 199-206
Abstract
Background: Direct aspiration thrombectomy is an increasingly utilized technique in endovascular stroke treatment that relies on vacuum and suction force to remove the clot. This report assesses the pressures and forces generated by different catheters and pumps. Methods: Vacuum pressures were measured using a vacuum gauge for several catheters (Stryker Catalyst6, Penumbra Ace 064, Medtronic Arc 061, and Penumbra 041) and pumps (Penumbra pump, 60-mL syringe, and the ASPIRE device). Suction forces were calculated based on pressure and catheter tip size (force = area × pressure). Vacuum pressures and forces were also assessed with a coaxial microcatheter (Rebar 18; inner diameter = 0.021 inches), mimicking a combined aspiration and stent retriever approach. Results: All catheters transmitted similar vacuum pressures, but suction force was proportionate to catheter tip area. Pump vacuum pressures were also similar, although the Penumbra pump (mean -25.63 inches Hg [inHg]) was slightly weaker than the syringe and the ASPIRE device (-27.04 and -27.58 inHg, respectively; p < 0.001, two-way ANOVA). A coaxial microcatheter lowered the mean vacuum pressure by only 1.2 inHg (p = 0.005), though it would theoretically create a very significant reduction in suction force if partially blocking the aspiration catheter tip area. Conclusions: All catheters transmit similar vacuum pressure, but the suction force on the clot is stronger with larger catheter tips. Coaxial microcatheters will have a minimal impact on suction force as long as they are proximal to the aspiration catheter tip. Currently available thrombectomy suction devices, including the Penumbra pump, the ASPIRE handheld pump, and the 60-mL syringe, all develop similar vacuum pressures.Keywords
This publication has 19 references indexed in Scilit:
- Thrombectomy in acute ischemic stroke: estimations of increasing demandsJournal of NeuroInterventional Surgery, 2016
- Long term experience using the ADAPT technique for the treatment of acute ischemic strokeJournal of NeuroInterventional Surgery, 2016
- Optimizing endovascular stroke treatment: removing the microcatheter before clot retrieval with stent-retrievers increases aspiration flowJournal of NeuroInterventional Surgery, 2016
- Comparison of clinical outcomes in patients with acute ischemic strokes treated with mechanical thrombectomy using either Solumbra or ADAPT techniquesJournal of NeuroInterventional Surgery, 2015
- 2015 American Heart Association/American Stroke Association Focused Update of the 2013 Guidelines for the Early Management of Patients With Acute Ischemic Stroke Regarding Endovascular TreatmentStroke, 2015
- Force and aspiration analysis of the ADAPT technique in acute ischemic stroke treatmentJournal of NeuroInterventional Surgery, 2015
- ADAPT FAST study: a direct aspiration first pass technique for acute stroke thrombectomyJournal of NeuroInterventional Surgery, 2014
- Distal aspiration with retrievable stent assisted thrombectomy for the treatment of acute ischemic strokeJournal of NeuroInterventional Surgery, 2014
- Initial clinical experience with the ADAPT technique: A direct aspiration first pass technique for stroke thrombectomyJournal of NeuroInterventional Surgery, 2013
- Hydrodynamic comparison of the Penumbra system and commonly available syringes in forced-suction thrombectomyJournal of NeuroInterventional Surgery, 2013