Simultaneous whole‐liver water T1 and T2 mapping with isotropic resolution during free‐breathing
Open Access
- 4 August 2024
- journal article
- research article
- Published by Wiley in NMR in Biomedicine
- Vol. 37 (12) , e5216
Abstract
Purpose: To develop and validate a data acquisition scheme combined with a motion‐resolved reconstruction and dictionary‐matching‐based parameter estimation to enable free‐breathing isotropic resolution self‐navigated whole‐liver simultaneous water‐specific ( ) and ( ) mapping for the characterization of diffuse and oncological liver diseases.Methods: The proposed data acquisition consists of a magnetization preparation pulse and a two‐echo gradient echo readout with a radial stack‐of‐stars trajectory, repeated with different preparations to achieve different and contrasts in a fixed acquisition time of 6 min. Regularized reconstruction was performed using self‐navigation to account for motion during the free‐breathing acquisition, followed by water–fat separation. Bloch simulations of the sequence were applied to optimize the sequence timing for insensitivity at 3 T, to correct for relaxation‐induced blurring, and to map and using a dictionary. The proposed method was validated on a water–fat phantom with varying relaxation properties and in 10 volunteers against imaging and spectroscopy reference values. The performance and robustness of the proposed method were evaluated in five patients with abdominal pathologies.Results: Simulations demonstrate good insensitivity of the proposed method in measuring and values. The proposed method produces co‐registered and maps with a good agreement with reference methods (phantom: ; ). The proposed and mapping exhibits good repeatability and can be robustly performed in patients with pathologies.Conclusions: The proposed method allows whole‐liver and quantification with high accuracy at isotropic resolution in a fixed acquisition time during free‐breathing.Keywords
Funding Information
- Philips
- International Graduate School of Science and Engineering
This publication has 69 references indexed in Scilit:
- Global Perspectives on Nonalcoholic Fatty Liver Disease and Nonalcoholic SteatohepatitisHepatology, 2019
- In vivo triglyceride composition of abdominal adipose tissue measured by 1H MRS at 3TJournal of Magnetic Resonance Imaging, 2016
- Influence of fat on liver T1 measurements using modified Look–Locker inversion recovery (MOLLI) methods at 3TJournal of Magnetic Resonance Imaging, 2016
- Motion artifacts in MRI: A complex problem with many partial solutionsJournal of Magnetic Resonance Imaging, 2015
- Quantification of liver iron with MRI: State of the art and remaining challengesJournal of Magnetic Resonance Imaging, 2014
- Magnetic resonance elastography of liver: Technique, analysis, and clinical applicationsJournal of Magnetic Resonance Imaging, 2013
- Proton density fat‐fraction: A standardized mr‐based biomarker of tissue fat concentrationJournal of Magnetic Resonance Imaging, 2012
- Quantitative assessment of liver fat with magnetic resonance imaging and spectroscopyJournal of Magnetic Resonance Imaging, 2011
- Repeatability of T1‐quantification in dGEMRIC for three different acquisition techniques: Two‐dimensional inversion recovery, three‐dimensional look locker, and three‐dimensional variable flip angleJournal of Magnetic Resonance Imaging, 2010
- Differentiation of hepatocellular carcinoma and hepatic metastasis from cysts and hemangiomas with calculated T2 relaxation times and the T1/T2 relaxation times ratioJournal of Magnetic Resonance Imaging, 2006