Biomechanical Analysis of Patellar Tendon Repair With Knotless Suture Anchor Tape Versus Transosseous Suture
Open Access
- 1 October 2020
- journal article
- research article
- Published by SAGE Publications in Orthopaedic Journal of Sports Medicine
- Vol. 8 (10)
Abstract
Background: Patellar tendon ruptures have routinely been repaired with transosseous suture tunnels. The use of knotless suture anchors for repair has been suggested as an alternative. Purpose: To compare the load to failure and gap formation of patellar tendon repair at the inferior pole of the patella with knotless suture anchor tape versus transosseous sutures. A secondary objective was to investigate whether either technique shows an association between bone density and load to failure. Study Design: Controlled laboratory study. Methods: A total of 20 human tibias with attached patellar and quadriceps tendons were sharply incised at the bone-tendon junction at the inferior pole of the patella. A total of 10 tendons were repaired using 2 knotless suture anchors in the inferior pole of the patella and a single suture tape with 2 core sutures. The other 10 tendons were repaired using No. 2 suture passed through 3 transosseous tunnels. A distracting force was then applied through the suture in the quadriceps tendon. Gap distance through load cycling at the repair site and maximum load at repair failure were then measured. Bone density was measured using computed tomography scanning. Results: No difference was found in the mean load to failure of knotless patellar tendon repair versus transosseous suture repair (367.6 ± 112.2 vs 433.9 ± 99 N, respectively; P = .12). After 250 cycles, the mean repair site gap distance was 0.85 ± 0.45 mm for the knotless patellar tendon repair versus 2.94 ± 2.03 mm for the transosseous suture repair ( P = .03). A small correlation, although not statistically significant, was found between bone density and load to failure for the knotless tape repair ( R 2 = 0.228; P = .66). No correlation was found between bone density and load to failure for the transosseous repair ( R 2 = 0.086; P = .83). Conclusion: Suture tape repair with knotless anchors for repair of patellar tendon rupture has comparable load to failure with less gap formation than transosseous suture repair. There is a small correlation between bone density and failure load for knotless anchor repair, which may benefit from further investigation. Clinical Relevance: Using knotless suture anchors for patellar tendon rupture repair would allow for a smaller incision, less dissection, and likely shorter operating time.Keywords
This publication has 22 references indexed in Scilit:
- Biomechanical Evaluation of Suture Anchor Versus Transosseous Tunnel Quadriceps Tendon Repair TechniquesArthroscopy, 2016
- High‐Tensile Strength Tape Versus High‐Tensile Strength Suture: A Biomechanical StudyArthroscopy, 2015
- Acute traumatic patellar tendon rupture: Early and late results of surgical treatment of 38 casesOrthopaedics & Traumatology: Surgery & Research, 2015
- Biomechanical Properties of Suture Anchor Repair Compared With Transosseous Sutures in Patellar Tendon RupturesThe American Journal of Sports Medicine, 2013
- EPIDEMIOLOGICAL STUDY ON TENDON RUPTURES OF THE KNEE EXTENSOR MECHANISM AT A LEVEL 1 HOSPITAL.Revista Brasileira de Ortopedia (English Edition), 2012
- Estudo epidemiológico das rupturas tendinosas do mecanismo extensor do joelho em um hospital de nível IRevista Brasileira de Ortopedia, 2012
- The epidemiology of musculoskeletal tendinous and ligamentous injuriesInjury, 2008
- Repair of Patellar Tendon Rupture With Suture AnchorsArthroscopy, 2007
- The Use of Suture Anchors in Repair of the Ruptured Patellar TendonThe American Journal of Sports Medicine, 2006
- Bilateral patellar tendon rupture without predisposing systemic disease or steroid use: a case report and review of the literatureArchives of orthopaedic and trauma surgery, 2005