Hydrodynamic function of tilting disc prostheses and bileaflet valves in double valve replacement

Authors:
  • ACA Campbell
    A Campbell
  • GRG Richard
    G Richard
  • JLJ Laas
    J Laas

Abstract
OBJECTIVE: To determine the energy loss attributable to prosthetic valvesize and design in double valve replacement, energy consumption of mitralvalves (size #25 to #29), of two different designs (Bjork Shiley tiltingdisc and Carbomedics bileaflet valves), in combination with a small (#21)and large sized (#27) aortic prosthesis, were analyzed in a flow simulator.METHODS: A pulsatile flow simulator was used to reproduce physiologicalventricular dynamics and to match the input and output impedances of thehuman left ventricle. Hydrodynamic performance was determined as strokework, closing work and leak work for each combination of valves at low flow(5 liters per minute) and high flow (9 liters per minute) conditions.RESULTS: At low flow no decrease of energy requirement was found with theuse of a mitral valve larger than #25. At high flow the #27 and #29 mitralvalves required less energy compared to a #25 mitral valve, in combinationwith a large aortic prosthesis. The #29 mitral prosthesis revealed similarresults as the #27. With the use of a large aortic prosthesis a remarkablereduction of total flow work was shown. These results were found in bothdesigns. In comparison of the two designs, tilting disc valves requiredmore energy for stroke and closure of the valve, although less energy forleakage. In total, energy requirements were higher for tilting disc valves.COMMENTS: A hydrodynamic advantage for the use of a mitral valve largerthan #25 was found only with the combination of a large aortic prosthesisand high flow. Hydrodynamic data favor the use of bileaflet prosthesisespecially for a patient who is expected to exercise.