Reducing Nonlinear Errors in Wearable Biomagnetic Measurements Enabled by SERF Atomic Magnetometers Through Real-Time Suppression of Triaxial DC Interferences
- 9 January 2025
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Instrumentation and Measurement
- Vol. 74 (00189456) , 1-11
Abstract
Aiming at the issue of substantial nonlinear error arising from triaxial DC magnetic field interferences in wearable and mobile measurement scenarios, a triaxial magnetic fields response model of single-beam atomic SERF magnetometer under non-zero magnetic condition is established, revealing that the effect of triaxial DC magnetic field interferences on the detected AC responses exhibits the nonlinear feature, while their impact on the DC response manifests as the interval-linear characteristic. Consequently, a closed-loop control system for mixed AC and DC magnetic fields is proposed. By leveraging the DC response within the feedback signal, a localized linearization control strategy is implemented across the triaxial magnetic fields, enabling direct and real-time suppression of DC interferences, and subsequently reducing their nonlinear impact on the measured AC response. Among them, the nonlinear impact caused by x-axis DC magnetic field interference decreases from 48.12% in the open-loop configuration to 0.01%. Additionally, it takes less time to suppress the interferences to below 2%, shortened from 4.7 s to 32 ms.Keywords
Funding Information
- National Science Fund for Distinguished Young Scholars (62225102)
- 2022 Industrial Technology Basic Public Service Platform Project (2022-189-181)
- National Natural Science Foundation of China (62073014)
This publication has 37 references indexed in Scilit:
- Analysis and suppression of magnetic field interference for SERF atomic magnetometerMeasurement, 2024
- Accurate and fast measurement of alkali-metal and noble-gas atoms spin polarizability based on frequency response in SERF co-magnetometerMeasurement, 2023
- Research on the anti-interference ability of optically pumped magnetometerMeasurement, 2023
- Enabling ambulatory movement in wearable magnetoencephalography with matrix coil active magnetic shieldingNeuroImage, 2023
- Magnetic field interference suppression for minimized SERF atomic magnetometerSensors and Actuators A: Physical, 2023
- Interference suppression techniques for OPM-based MEG: Opportunities and challengesNeuroImage, 2022
- Cross-Axis projection error in optically pumped magnetometers and its implication for magnetoencephalography systemsNeuroImage, 2021
- Using OPMs to measure neural activity in standing, mobile participantsNeuroImage, 2021
- Optically pumped magnetometers: From quantum origins to multi-channel magnetoencephalographyNeuroImage, 2019
- On-scalp MEG system utilizing an actively shielded array of optically-pumped magnetometersNeuroImage, 2019