Non-orthogonal one-step calibration method for robotized transcranial magnetic stimulation
Open Access
- 1 October 2018
- journal article
- research article
- Published by Springer Nature in BioMedical Engineering OnLine
- Vol. 17 (1) , 137
Abstract
Robotized transcranial magnetic stimulation (TMS) combines the benefits of neuro-navigation with automation and provides a precision brain stimulation method. Since the coil will normally remain unmounted between different clinical uses, hand/eye calibration and coil calibration are required before each experiment. Today, these two steps are still separate: hand/eye calibration is performed using methods proposed by Tsai/Lenz or Floris Ernst, and then the coil calibration is carried out based on the traditional TMS experimental step. The process is complex and time-consuming, and traditional coil calibration using a handheld probe is susceptible to greater calibration error. A novel one-step calibration method has been developed to confirm hand/eye and coil calibration results by formulating a matrix equation system and estimating its solution. Hand/eye calibration and coil calibration are performed to confirm the pose relationships of the marker/end effector ‘X’, probe/end effector ‘Y’, and robot/world ‘Z’. First, the coil is fixed on the end effector of the robot. During the one-step calibration process, a marker is mounted on the top of the coil and a calibration probe is fixed at the actual effective position of the coil. Next, the robot end effector is moved to a series of random positions ‘A’, the tracking data of marker ‘B’ and probe ‘C’ is obtained correspondingly. Then, a matrix equation system AX = ZB and AY = ZC can be acquired, and it is computed using a least-squares approach. Finally, the calibration probe is removed after calibration, while the marker remains fixed to the coil during the TMS experiment. The methods were evaluated based on simulation data and on experimental data from an optical tracking device. We compared our methods with two classical methods: the QR24 method proposed by Floris Ernst and the handheld coil calibration method. The new methods outperform the QR24 method in the aspect of translational accuracy and performs similarly in the aspect of rotational accuracy, the total translational error decreased more than fifty percent. The new approach also outperforms traditional handheld coil calibration of navigated TMS systems, the total translational error decreased three- to fourfold, and the rotational error decreased six- to eightfold. Furthermore, the convergence speed is improved 16- to 27-fold for the new algorithms. These results suggest that the new method can be used for hand/eye and coil calibration of a robotized TMS system. Two complex steps can be simplified using a least-squares approach.Keywords
Funding Information
- National Natural Science Foundation of China (81772003)
- CAMS Initiative for Innovative Medicine (CAMS-I2M 2016-I2M-1004)
- PUMC Youth Fund and the Fundamental Research Funds for the Central Universities (2017320025)
This publication has 33 references indexed in Scilit:
- P 226. Robot guided positioning of a magnetic coil for rTMS over the cortex in patients with cerebral lesionsClinical Neurophysiology, 2013
- Non‐orthogonal tool/flange and robot/world calibrationThe International Journal of Medical Robotics and Computer Assisted Surgery, 2012
- Robust real‐time robot–world calibration for robotized transcranial magnetic stimulationThe International Journal of Medical Robotics and Computer Assisted Surgery, 2011
- Robot-assisted image-guided transcranial magnetic stimulation for somatotopic mapping of the motor cortex: a clinical pilot studyActa Neurochirurgica, 2009
- Transcranial Magnetic Stimulation: A PrimerNeuron, 2007
- Transcranial magnetic stimulation and the challenge of coil placement: A comparison of conventional and stereotaxic neuronavigational strategiesHuman Brain Mapping, 2007
- Transcranial Magnetic Stimulation in the Treatment of DepressionAmerican Journal of Psychiatry, 2003
- Extrinsic calibration of a vision sensor mounted on a robotIEEE Transactions on Robotics and Automation, 1992
- A new technique for fully autonomous and efficient 3D robotics hand/eye calibrationIEEE Transactions on Robotics and Automation, 1989
- Closed-form solution of absolute orientation using unit quaternionsJournal of the Optical Society of America A, 1987