DEVELOPMENT AND VALIDATION OF RP-HPLC METHOD FOR QUANTITATION OF AMOXICILLIN TRIHYDRATE IN BULK AND PHARMACEUTICAL FORMULATIONS USING BOX-BEHNKEN EXPERIMENTAL DESIGN
- 7 February 2012
- journal article
- research article
- Published by Taylor & Francis in Journal of Liquid Chromatography & Related Technologies
- Vol. 35 (3) , 393-406
Abstract
A rapid and sensitive novel reversed phase-HPLC method with UV detection at 229 nm was developed for routine analysis of amoxicillin trihydrate in bulk drugs and pharmaceutical formulations (Damoxy and Amoxil). Chromatographic separation was performed on a Capacel Pak C18 type MG column using isocratic elution of mobile phase containing mixture of phosphate buffer and methanol (50:50, %v/v) with flow rate at 1.0 ml/min and pH 3.0. The method was validated for linearity (r2 = 0.999), accuracy, precision, sensitivity, system suitability, selectivity, and robustness testing. A 32-factorial design was employed utilizing box-behnken experimental design for the validation of robustness using three factors such as composition of mobile phase (X1), flow rate (X2), pH (X3); while peak area (Y1) and retention time (Y2) were used as responses. System showed minor change in the mobile phase composition and flow rate affects the responses produced, while pH has no significant effect on the responses. Finally the developed experimental design was validated by statistical analysis using ANOVA showed that the proposed method was simple, sensitive, and highly robust for routine analysis of the amoxicillin trihydrate.Keywords
This publication has 15 references indexed in Scilit:
- RP-HPLC method for analysis of related substances in amoxicillin drug substanceActa Chromatographica, 2009
- Application and validation of a LC/fluorescence method for the determination of amoxicillin in sheep serum and tissue cage fluidJournal of Pharmaceutical and Biomedical Analysis, 2009
- Simultaneous determination of amoxicillin and ambroxol in human plasma by LC–MS/MS: Validation and application to pharmacokinetic studyJournal of Pharmaceutical and Biomedical Analysis, 2008
- Simultaneous determination of amoxicillin and clavulanic acid in human plasma by isocratic reversed-phase HPLC using UV detectionJournal of Pharmaceutical and Biomedical Analysis, 2007
- A sensitive assay of amoxicillin in mouse serum and broncho-alveolar lavage fluid by liquid–liquid extraction and reversed-phase HPLCJournal of Pharmaceutical and Biomedical Analysis, 2005
- Rapid determination of amoxicillin in premixes by HPLCJournal of Pharmaceutical and Biomedical Analysis, 2005
- Simultaneous determination of amoxicillin and clavulanic acid in human plasma by HPLC–ESI mass spectrometryJournal of Chromatography B, 2004
- Impact of acid secretion, gastritis, and mucus thickness on gastric transfer of antibiotics in ratsGut, 2002
- Measurement of amoxicillin in plasma and gastric samples using high-performance liquid chromatography with fluorimetric detectionJournal of Chromatography B, 2002
- Quantitative determination of amoxicillin and its decomposition products by high-performance liquid chromatographyJournal of Chromatography A, 1985