Conductance and F45° in the detection and differential diagnosis of middle ear pathologies in comparison to resonant frequency

Authors:
  • MEMegha
    Megha

Abstract
Purpose: This study aimed to compare the clinical utility of two measures of multifrequency tympanometry – F45 and conductance – at resonant frequency (RF) with that of RF. Materials and Methods: RF, F45°, and conductance at RF (ΔG) were determined in 109 participants with normal and 84 participants with pathological middle ears, all with either “As” or “Ad” tympanograms. The measures were compared between normal and pathological ears (otosclerosis, otitis media, ossicular chain discontinuity, and monomeric tympanic membrane). The sensitivity and specificity of the measures in detecting different middle ear pathologies were determined. The ability of the measures to differentiate the middle ear pathologies was derived through discriminant analysis. Results: There was a significant difference across middle ear pathologies in RF, F45°, and ΔG. RF showed the highest sensitivity in detecting otitis media, whereas ΔG showed the highest sensitivity in detecting otosclerosis. The measures could account for 94.8% of the variance in discriminant function. Conclusion: Conductance-based parameters can effectively detect and differentiate middle ear pathologies. Conductance at RF is better than RF in detecting otosclerosis. The measures can differentiate the pathologies even when the total admittance, as measured in single-frequency tympanometry, fail to differentiate.

This publication has 32 references indexed in Scilit: