Magnetic Coil Stimulation of the Spinal cord in the Dog Effect of Removal of Bony Structure on Eddy Current

Authors:

Abstract
Magnetic coil stimulation applied to the spine excites the lumosacral roots rather than the spinal cord itself, To evaluate the mechanism responsible for this phenomenon, the effect of volume conduction on the activation of neural elements were studied. In ten intact dogs compound muscle action potentials could be recorded in the soleus muscle only when the magnetic coil was placed over the cauda equina. After laminectomy and pediculotomy, compound muscle action potentials also could be elicited when the magnetic coil was placed over the spinal cord. These findings suggest that the bony structure surrounding the spinal cord interferas with the spread of magnetically induced eddy currents to the spinal cord.