Reducing RF‐related heating of cardiac pacemaker leads in MRI: Implementation and experimental verification of practical design changes
- 1 March 2012
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 68 (6) , 1963-1972
Abstract
There are serious concerns regarding safety when performing magnetic resonance imaging in patients with implanted conductive medical devices, such as cardiac pacemakers, and associated leads, as severe incidents have occurred in the past. In this study, several approaches for altering an implant's lead design were systematically developed and evaluated to enhance the safety of implanted medical devices in a magnetic resonance imaging environment. The individual impact of each design change on radiofrequency heating was then systematically investigated in functional lead prototypes at 1.5 T. Radiofrequency‐induced heating could be successfully reduced by three basic changes in conventional pacemaker lead design: (1) increasing the lead tip area, (2) increasing the lead conductor resistance, and (3) increasing outer lead insulation conductivity. The findings show that radiofrequency energy pickup in magnetic resonance imaging can be reduced and, therefore, patient safety can be improved with dedicated construction changes according to a “safe by design” strategy. Incorporation of the described alterations into implantable medical devices such as pacemaker leads can be used to help achieve favorable risk‐benefit‐ratios when performing magnetic resonance imaging in the respective patient group. Magn Reson Med, 2012.Keywords
Funding Information
- Bayerische Forschungsstiftung
This publication has 25 references indexed in Scilit:
- Magnetic resonance imaging in patients with a pacemaker system designed for the magnetic resonance environmentHeart Rhythm, 2011
- Safety of magnetic resonance imaging of patients with a new Medtronic EnRhythm MRI SureScan pacing system: clinical study designTrials, 2008
- Spatial distribution of RF‐induced E‐fields and implant heating in MRIMagnetic Resonance in Medicine, 2008
- Safety of Magnetic Resonance Imaging in Patients With Cardiovascular DevicesCirculation, 2007
- Strategy for Safe Performance of Extrathoracic Magnetic Resonance Imaging at 1.5 Tesla in the Presence of Cardiac Pacemakers in Non–Pacemaker-Dependent PatientsCirculation, 2006
- Transmission line for improved RF safety of interventional devicesMagnetic Resonance in Medicine, 2005
- In vivo heating of pacemaker leads during magnetic resonance imagingEuropean Heart Journal, 2004
- Modern Pacemaker and Implantable Cardioverter/Defibrillator Systems Can Be Magnetic Resonance Imaging SafeCirculation, 2004
- Magnetic resonance imaging and cardiac pacemaker safety at 1.5-TeslaJACC, 2004
- Effects of magnetic resonance imaging on cardiac pacemakers and electrodesAmerican Heart Journal, 1997