Abstract
Tumor motion induced by respiration presents a challenge to the reliable delivery of conformal radiation treatments. Real-time motion compensation represents the technologically most challenging clinical solution but has the potential to overcome the limitations of existing methods. The performance of a real-time couch-based motion compensation system is mainly dependent on two aspects: the ability to infer the internal anatomical position and the performance of the feedback control system. In this paper, we propose two novel methods for the two aspects respectively, and then combine the proposed methods into one system. To accurately estimate the internal tumor position, we present partial-least squares (PLS) regression to predict the position of the diaphragm using skin-based motion surrogates. Four radio-opaque markers were placed on the abdomen of patients who underwent fluoroscopic imaging of the diaphragm. The coordinates of the markers served as input variables and the position of the diaphragm served as the output variable. PLS resulted in lower prediction errors compared with standard multiple linear regression (MLR). The performance of the feedback control system depends on the system dynamics and dead time (delay between the initiation and execution of the control action). While the dynamics of the system can be inverted in a feedback control system, the dead time cannot be inverted. To overcome the dead time of the system, we propose a predictive feedback control system by incorporating forward prediction using least-mean-square (LMS) and recursive least square (RLS) filtering into the couch-based control system. Motion data were obtained using a skin-based marker. The proposed predictive feedback control system was benchmarked against pure feedback control (no forward prediction) and resulted in a significant performance gain. Finally, we combined the PLS inference model and the predictive feedback control to evaluate the overall performance of the feedback control system. Our results show that, with the tumor motion unknown but inferred by skin-based markers through the PLS model, the predictive feedback control system was able to effectively compensate intra-fraction motion.For more information on this article, see medicalphysicsweb.orgKeywords
This publication has 34 references indexed in Scilit:
- An analysis of the treatment couch and control system dynamics for respiration‐induced motion compensationMedical Physics, 2006
- Geometric accuracy of a real-time target tracking system with dynamic multileaf collimator tracking systemInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2006
- Insertion and fixation of fiducial markers for setup and tracking of lung tumors in radiotherapyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2005
- On using an adaptive neural network to predict lung tumor motion during respiration for radiotherapy applicationsMedical Physics, 2005
- Real-time intra-fraction-motion tracking using the treatment couch: a feasibility studyPhysics in Medicine & Biology, 2005
- Factors Associated With Pneumothorax and Pneumothorax Requiring Treatment after Percutaneous Lung Biopsy in 443 Consecutive PatientsJournal of Vascular and Interventional Radiology, 2004
- Robotics and Radiosurgery – The CyberknifeStereotactic and Functional Neurosurgery, 2001
- Motion adaptive x-ray therapy: a feasibility studyPhysics in Medicine & Biology, 2000
- Correction for geometric image distortion in the x‐ray imaging chain: Local technique versus global techniqueMedical Physics, 1999
- SIMPLS: An alternative approach to partial least squares regressionChemometrics and Intelligent Laboratory Systems, 1993