The dose from Compton backscatter screening
Authors:
- 9 November 2010
- journal article
- research article
- Published by Oxford University Press (OUP) in Radiation Protection Dosimetry
- Vol. 145 (1) , 75-81
Abstract
Systems based on the detection of Compton backscattered X rays have been deployed for screening personnel for weapons and explosives. Similar principles are used for screening vehicles at border-crossing points. Based on well-established scattering cross sections and absorption coefficients in conjunction with reasonable estimates of the image contrast and resolution, the entrance skin dose and the dose at a depth of 1 cm can be calculated. The effective dose can be estimated using the same conversion coefficients as used to convert exposure measurements to the effective dose. It is shown that the effective dose is highly dependent on image resolution (i.e. pixel size).The effective doses for personnel screening systems are unlikely to be in compliance with the American National Standards Institute standard NS 43.17 unless the pixel sizes are >4 mm. Nevertheless, calculated effective doses are well below doses associated with health effects.Keywords
This publication has 2 references indexed in Scilit:
- Determination of ambient and personal dose equivalent for personnel and cargo security screeningRadiation Protection Dosimetry, 2006
- PENELOPE: An algorithm for Monte Carlo simulation of the penetration and energy loss of electrons and positrons in matterNuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms, 1995