Comparison of hematopoietic progenitor cell collection using different inlet flow rates with the Fenwal Amicus
- 12 January 2022
- journal article
- research article
- Published by Wiley in Journal of Clinical Apheresis
- Vol. 37 (3) , 206-216
Abstract
Purpose: We have used a hematopoietic progenitor cell (HPC) algorithm (standard [STD]) that restricted the inlet flow rate to 65 mL/min for peripheral white blood cell count (PWBC) >35 × 109/L (STD). In this study, we evaluated a technique that allows 85 mL/min, regardless of the PWBC count (high). For patients with PWBC >35 × 109/L, a prospective, randomized comparison of the high flow rate vs the STD PWBC‐based flow rate (65 mL/min) was performed, comparing CD34+ and lymphocyte yields, collection efficiencies (CE1), mononuclear cells (MNC), and granulocytes, red blood cell (RBC), and platelet content.Methods: The Fenwal Amicus version 4.5 with a heparinized ACD‐A anticoagulant (AC) delivered at a 26:1 AC ratio was used. Paired comparisons between high and STD techniques were assessed with Wilcoxon signed rank tests, with P < .05 considered significant. Data are summarized as medians.Results: Forty patient pairs (autologous) were compared. Diagnoses included primarily multiple myeloma (60%) and lymphoma (37.5%). High had significantly higher median average inlet rates (69 vs 55 mL/min), whole blood processed (20 vs 16 L), and cycles (15 vs 14) than STD. There were no significant differences in pre‐procedure counts. Collection contents were (high/STD): 306/328 × 106 CD34+ cells, 48/59% CD34+ CE1 (significant), 0.2/0.2 × 109/kg lymphocytes, 45/57% lymphocyte CE1, 63/59 × 109 WBC, 15/16 × 109 granulocytes, and 1.9/1.7 × 1011 platelets.Conclusions: The simpler, standardized high flow technique did not significantly increase or decrease CD34+ cells or lymphocyte yields, but did significantly decrease CD34+ CE1. The effects on cross‐cellular content were minimal and not clinically significant.Keywords
This publication has 21 references indexed in Scilit:
- HPC‐A dose prediction on the optia® cell separator based on a benchmark CE2 collection efficiency: Promoting clinical efficiency, minimizing toxicity, and allowing quality controlJournal of Clinical Apheresis, 2015
- Infused Autograft Lymphocyte to Monocyte Ratio and Survival in Diffuse Large B Cell LymphomaTransplantation and Cellular Therapy, 2014
- Evaluation of the spectra Optia® mononuclear cell collection procedure in multiple myeloma patientsJournal of Clinical Apheresis, 2014
- A single center comparison between three different apheresis systems for autologous and allogeneic stem cell collectionsTransfusion and Apheresis Science, 2013
- Optimization of MNC collection efficiency depending on patient WBC precount with AmicusTransfusion and Apheresis Science, 2013
- Comparing peripheral blood stem cell collection using the COBE Spectra, Haemonetics MCS+, and Baxter AmicusTransfusion and Apheresis Science, 2012
- A multiple regression analysis on factors influencing haematopoietic progenitor cell collection for autologous transplantationTransfusion and Apheresis Science, 2012
- Manual color monitoring to optimize hematopoietic progenitor cell collection on the Fenwal AmicusJournal of Clinical Apheresis, 2011
- A prospective randomized trial of two popular mononuclear cell collection sets for autologous peripheral blood stem cell collection in multiple myelomaTransfusion, 2010
- Hematopoietic progenitor cell large volume leukapheresis (LVL) on the Fenwal Amicus blood separatorJournal of Clinical Apheresis, 2004