Changes in surface topography at the TKA backside articulation following in vivo service: a retrieval analysis
- 7 August 2014
- journal article
- knee
- Published by Wiley in Knee Surgery, Sports Traumatology, Arthroscopy
- Vol. 23 (12) , 3523-3531
Abstract
With the advent of modular total knee arthroplasty (TKA) systems, backside wear at the articulation between the ultra-high-molecular-weight-polyethylene (UHMWPE) component undersurface and the tibial baseplate has received increasing attention as a source of clinically significant polyethylene wear debris. The aim of this study was to investigate the reciprocating interface at the TKA undersurface articulation using profilometry after in vivo service. Our null hypothesis was that there would be no discernible pattern or relationship between the metal tibial baseplate and UHMWPE surface profile.Keywords
This publication has 24 references indexed in Scilit:
- Retrieval analysis of modular total knee replacements: Factors influencing backside surface damageThe Knee, 2012
- Second-Generation Locking Mechanisms and Ethylene Oxide Sterilization Reduce Tibial Insert Backside Damage in Total Knee ArthroplastyThe Journal of Arthroplasty, 2011
- Quantification of self-polishing in vivo from explanted metal-on-metal total hip replacementsTribology International, 2011
- Delamination wear on two retrieved polyethylene inserts after gamma sterilization in nitrogenThe Knee, 2011
- Material and surface factors influencing backside fretting wear in total knee replacement tibial componentsJournal of Biomechanics, 2010
- Tribological analysis of failed resurfacing hip prostheses and comparison with clinical dataProceedings of the Institution of Mechanical Engineers, Part J: Journal of Engineering Tribology, 2009
- Polyethylene Wear Is Influenced by Manufacturing Technique in Modular TKAClinical Orthopaedics and Related Research, 2008
- Backside Wear in Modern Total Knee DesignsHSS Journal®: The Musculoskeletal Journal of Hospital for Special Surgery, 2007
- The role of counterface imperfections in the wear of polyethyleneWear, 1987
- Retrieval analysis of total knee prostheses: A method and its application to 48 total condylar prosthesesJournal of Biomedical Materials Research, 1983