Abstract
Cochlear Implants (CIs) are medical implantable devices that can restore the sense of hearing in people with profound hearing loss. Clinical trials assessing speech intelligibility in CI users have found large inter-subject variability. One possibility to explain the variability is the individual differences in the interface created between electrodes of the CI and the auditory nerve. In order to understand the variability, models of the voltage distribution of the electrically stimulated cochlea may be useful. With this purpose in mind, we developed a parametric model that can be adapted to each CI user based on landmarks from individual cone beam computed tomography (CBCT) scans of the cochlea before and after implantation. The conductivity values of each cochlea compartment as well as the weighting factors of different grounding modes have been also parameterized. Simulations were performed modeling the cochlea and electrode positions of 12 CI users. Three models were compared with different levels of detail: A homogeneous model (HM), a non-patient specific model (NPSM) and a patient specific model (PSM). The model simulations were compared with voltage distribution measurements obtained from the backward telemetry of the 12 CI users. Results show that the PSM produces the lowest error when predicting individual voltage distributions. Given a patient specific geometry and electrode positions we show an example on how to optimize the parameters of the model and how to couple it to an auditory nerve model. The model here presented may help to understand speech performance variability and support the development of new sound coding strategies for CIs.Keywords
Funding Information
- Deutsche Forschungsgemeinschaft (Cluster of Excellence EXC 1077/1 “Hearing4all”.)
This publication has 36 references indexed in Scilit:
- Reducing current spread using current focusing in cochlear implant usersHearing Research, 2012
- A Longitudinal Study on Auditory Perception and Speech Intelligibility in Deaf Children Implanted Younger Than 18 Months in Comparison to Those Implanted at Later AgesOtology & Neurotology, 2010
- Signal Processing Strategies for Cochlear Implants Using Current SteeringEURASIP Journal on Advances in Signal Processing, 2009
- Effects of steroids and lubricants on electrical impedance and tissue response following cochlear implantationCochlear Implants International, 2007
- The Size of the Cochlea and Predictions of Insertion Depth Angles for Cochlear Implant ElectrodesAudiology and Neurotology, 2006
- Three-dimensional modeling and visualization of the cochlea on the InternetIEEE Transactions on Information Technology in Biomedicine, 2000
- Potential distributions and neural excitation patterns in a rotationally symmetric model of the electrically stimulated cochleaHearing Research, 1995
- Lumped-parameter model for in vivo cochlear stimulationIEEE Transactions on Biomedical Engineering, 1993
- Better speech recognition with cochlear implantsNature, 1991
- A cochlear frequency-position function for several species—29 years laterThe Journal of the Acoustical Society of America, 1990