Cross-reaction of POC-CCA urine test for detection of Schistosoma mekongi in Lao PDR: a cross-sectional study
Open Access
- 12 August 2020
- journal article
- research article
- Published by Springer Nature in Infectious Diseases of Poverty
- Vol. 9 (1) , 1-9
Abstract
Background: The point-of-care circulating cathodic antigen (POC-CCA) test is increasingly used as a rapid diagnostic method for Schistosoma mansoni infection. The test has good sensitivity, although false positive results have been reported among pregnant women and patients with urine infections and hematuria. We validated the POC-CCA test’s ability to diagnose Schistosoma mekongi infection in Lao People’s Democratic Republic (Lao PDR), where S. mekongi is endemic. Of particular interest was the test’s specificity and possible cross-reactivity with other helminth infections. Methods: We conducted a cross-sectional study of children and adults in the provinces of Champasack (Schistosoma mekongi and Opisthorchis viverrini endemic), Savannakhet (O. viverrini endemic) and Luang Prabang (soil-transmitted helminths endemic) between October 2018 and April 2019. POC-CCA and urine dipstick tests were administered to all study participants, while an additional pregnancy test was offered to women. Two stool samples were collected from participants and examined with a Kato-Katz test (two smears per stool). Logistic regression was used to associate potential confounding factors (predictors) with POC-CCA test results (outcome). Results: In S. mekongi-endemic Champasack, 11.5% (n = 366) and 0.5% (n = 2) of study participants had positive POC-CCA and Kato-Katz test results, respectively. Only one of the two Kato-Katz positive patients was also POC-CCA positive. In Champasack and Luang Prabang, where S. mekongi is not endemic, the POC-CCA test yielded (presumably) false positive results for 6.0% (n = 22) and 2.5% (n = 9) of study participants, respectively, while all of the Kato-Katz tests were negative. POC-CCA positive test results were significantly associated with O. viverrini infection (1.69, 95% confidence interval (CI): 1.02–2.77, P = 0.042), increased leukocytes (adjusted Odds Ratio (aOR) = 1.58, 95% CI: 1.15–2.17, P = 0.005) and hematuria (aOR = 1.50, 95% CI: 1.07–2.10, P = 0.019) if the observed trace was counted as a positive test result. Two pregnant women from Champasack province had POC-CCA positive tests. Conclusions: We observed a cross-reaction between the POC-CCA test and O. viverrini infection. To some extent, we can confirm previous observations asserting that POC-CCA provides false positive results among patients with urinary tract infections and hematuria. In S. mekongi-endemic areas, POC-CCA can be applied cautiously for surveillance purposes, keeping in mind the considerable risk of false positive results and its unknown sensitivity.Keywords
This publication has 22 references indexed in Scilit:
- Performance of POC-CCA® in diagnosis of schistosomiasis mansoni in individuals with low parasite burdenRevista da Sociedade Brasileira de Medicina Tropical, 2016
- Validation of a Point-of-Care Circulating Cathodic Antigen Urine Cassette Test for Schistosoma mansoni Diagnosis in the Sahel, and Potential Cross-Reaction in PregnancyThe American Journal of Tropical Medicine and Hygiene, 2016
- A Five-Country Evaluation of a Point-of-Care Circulating Cathodic Antigen Urine Assay for the Prevalence of Schistosoma mansoniThe American Journal of Tropical Medicine and Hygiene, 2013
- Efficacy of Praziquantel against Schistosoma mekongi and Opisthorchis viverrini: A Randomized, Single-Blinded Dose-Comparison TrialPLOS Neglected Tropical Diseases, 2012
- Helminth and Intestinal Protozoa Infections, Multiparasitism and Risk Factors in Champasack Province, Lao People's Democratic RepublicPLOS Neglected Tropical Diseases, 2011
- Diversity of human intestinal helminthiasis in Lao PDRTransactions of the Royal Society of Tropical Medicine and Hygiene, 2009
- Evaluation of circulating cathodic antigen (CCA) strip for diagnosis of urinary schistosomiasis in Hassoba school children, Afar, EthiopiaParasite, 2008
- Up-Converting Phosphor Technology-Based Lateral Flow Assay for Detection of Schistosoma Circulating Anodic Antigen in SerumJournal of Clinical Microbiology, 2008
- Schistosomiasis mekongi: from discovery to controlParasitology International, 2004
- Epidemiology and control of mekongi schistosomiasisActa Tropica, 2002