Effective Doses in Children: Association With Common Complex Imaging Techniques Used During Interventional Radiology Procedures
- 1 December 2014
- journal article
- research article
- Published by American Roentgen Ray Society in American Journal of Roentgenology
- Vol. 203 (6) , 1336-1344
Abstract
OBJECTIVE. The purpose of this study was to determine the range of effective doses associated with imaging techniques used during interventional radiology procedures on children. MATERIALS AND METHODS. A pediatric phantom set (1, 5, and 10 years) coupled with high-sensitivity metal oxide semiconductor field effect transistor (MOSFET) dosimeters was used to calculate effective doses. Twenty MOSFETs were inserted into each phantom at radiosensitive organ locations. The phantoms were exposed to mock head, chest, and abdominal interventional radiology procedures performed with different geometries and magnifications. Fluoroscopy, digital subtraction angiography (DSA), and spin angiography were simulated on each phantom. Road mapping was conducted only on the 5-year-old phantom. International Commission on Radiological Protection publication 103 tissue weights were applied to the organ doses recorded with the MOSFETs to determine effective dose. For easy application to clinical cases, doses were normalized per minute of fluoroscopy and per 10 frames of DSA or spin angiography. RESULTS. Effective doses from DSA, angiography, and fluoroscopy were higher for younger ages because of magnification use and were largest for abdominal procedures. DSA of the head, chest, and abdomen (normalized per 10 frames) imparted doses 2-3 times as high as corresponding doses per minute of fluoroscopy while all other factors remained unchanged (age, projection, collimation, magnification). Three to five frames of DSA imparted an effective dose equal to doses from 1 minute of fluoroscopy. Doses from spin angiography were almost one-half the doses received from an equivalent number of frames of DSA. CONCLUSION. Patient effective doses during interventional procedures vary substantially depending on procedure type but tend to be higher because of magnification use in younger children and higher in the abdomen.Keywords
This publication has 32 references indexed in Scilit:
- Comparison of indirect radiation dose estimates with directly measured radiation dose for patients and operators during complex endovascular proceduresJournal of Vascular Surgery, 2011
- Effective dose analysis of three-dimensional rotational angiography during catheter ablation proceduresPhysics in Medicine & Biology, 2010
- EXTRAPOLATING RADIATION-INDUCED CANCER RISKS FROM LOW DOSES TO VERY LOW DOSESHealth Physics, 2009
- Managing patient dose in interventional cardiologyCatheterization and Cardiovascular Interventions, 2007
- Validation of Metal Oxide Semiconductor Field Effect Transistor Technology for Organ Dose Assessment During CT: Comparison with Thermoluminescent DosimetryAmerican Journal of Roentgenology, 2007
- Pediatric interventional radiography equipment: safety considerationsPediatric Radiology, 2006
- Practice of ALARA in the pediatric interventional suitePediatric Radiology, 2006
- SKIN ENTRANCE RADIATION DOSE IN AN INTERVENTIONAL RADIOLOGY PROCEDUREHealth Physics, 2006
- Estimating cancer risks from pediatric CT: going from the qualitative to the quantitativePediatric Radiology, 2002
- IntroductionAnnals of the ICRP, 1991