Modalities of Invasive Arterial Pressure Monitoring in Critically Ill Patients
Authors:
- Karine GrittiKGKarine Gritti
- Cécile CarréCCCécile Carré
- Nadège FleuryNFNadège Fleury
- Annie LangALAnnie Lang
- Josette Courau-CourtoisJCJosette Courau-Courtois
- Jean-Pierre BedosJBJean-Pierre Bedos
- Stephane LegrielSLStephane Legriel
Open Access
- 1 September 2015
- journal article
- research article
- Published by Wolters Kluwer Health in Medicine
- Vol. 94 (39) , e1557
Abstract
Few studies assessed modalities of invasive arterial pressure monitoring (IAPM). We evaluated effects on measured values of various combinations of transducer level, catheter access site, and patient position. Prospective observational study in consecutive adults admitted to a French intensive care unit in 2009 to 2011 and fulfilling our inclusion criteria. Four combinations (B-E) of transducer level, catheter access site, and patient position were compared with a reference combination (A) (A: patient supine with all catheters in the same plane and a single transducer level (M) for zero point reference (Z) aligned on the phlebostatic axis; B: 45° head-of-bed elevation with M and Z aligned on the phlebostatic axis; C: 45° head-of-bed elevation with M aligned on the catheter access site and Z on the phlebostatic axis; D: 45° head-of-bed elevation with M and Z aligned on the catheter access site; and E: 45° head-of-bed elevation with M aligned on the phlebostatic axis and Z on the catheter access site). We included 103 patients, 68 men and 35 women, with a median age of 69 years (interquartile range [IQR], 56-78); at inclusion, 91 (88.3%) received mechanical ventilation, 45 (43.7%) catecholamines, and 66 (64.1%) sedation. The IAPM access site was femoral in 49 (47.6%) and radial in 54 (52.4%) patients, with 62 of 103 (60.2%) catheters on the right side. Measured absolute mean arterial pressure values were significantly higher with 3 study combinations (C-E) than with the reference combination (A). After adjustment, the differences versus A (median, 83 [IQR, 74-92] mm Hg) remained significant for D (median, 91 [IQR, 85-100] mm Hg, P < 0.001) and E (median, 88 [IQR, 77-99] mm Hg, P < 0.001). The difference versus A was not significant for B (median, 85 [IQR, 76-94] mm Hg, P = 0.21) or C (median, 90 [IQR, 84-100] mm Hg, P = 0.006). Several modalities used for zeroing and/or transducer leveling during IAPM may result in statistically and clinically significant overestimation of measured mean arterial pressure values. For patients in the 45° head-of-bed elevation position, aligning the Z on the phlebostatic axis provides values that are not significantly different from those obtained using the reference supine modality.Keywords
This publication has 25 references indexed in Scilit:
- A Comparison of Clinical and Research Practices in Measuring Cerebral Perfusion PressureAnesthesia & Analgesia, 2013
- Evidence-Based Practice Habits: Transforming Research Into Bedside PracticeCritical Care Nurse, 2009
- Critical care in the emergency department: monitoring the critically ill patientEmergency Medicine Journal, 2006
- Common errors in clinical measurementAnaesthesia & Intensive Care Medicine, 2005
- Changes in Arterial Pressure during Mechanical VentilationAnesthesiology, 2005
- Recommendations for Blood Pressure Measurement in Humans and Experimental AnimalsCirculation, 2005
- Inter-arm differences in blood pressure: when are they clinically significant?Journal Of Hypertension, 2002
- The Prevention of Ventilator-Associated PneumoniaNew England Journal of Medicine, 1999
- Comparison of direct blood pressure measurement at the radial and dorsalis pedis arteries during surgery in the horizontal and reverse Trendelenburg positionsAnaesthesia, 1995
- A new Simplified Acute Physiology Score (SAPS II) based on a European/North American multicenter studyJAMA, 1993