Model‐based image reconstruction for four‐dimensional PET
Authors:
- 21 April 2006
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 33 (5) , 1288-1298
Abstract
Positron emission tonography (PET) is useful in diagnosis and radiation treatment planning for a variety of cancers. For patients with cancers in thoracic or upper abdominal region, the respiratory motion produces large distortions in the tumor shape and size, affecting the accuracy in both diagnosis and treatment. Four‐dimensional (4D) (gated) PET aims to reduce the motion artifacts and to provide accurate measurement of the tumor volume and the tracer concentration. A major issue in 4D PET is the lack of statistics. Since the collected photons are divided into several frames in the 4D PET scan, the quality of each reconstructed frame degrades as the number of frames increases. The increased noise in each frame heavily degrades the quantitative accuracy of the PET imaging. In this work, we propose a method to enhance the performance of 4D PET by developing a new technique of 4D PET reconstruction with incorporation of an organ motion model derived from 4D‐CT images. The method is based on the well‐known maximum‐likelihood expectation‐maximization (ML‐EM) algorithm. During the processes of forward‐ and backward‐projection in the ML‐EM iterations, all projection data acquired at different phases are combined together to update the emission map with the aid of deformable model, the statistics is therefore greatly improved. The proposed algorithm was first evaluated with computer simulations using a mathematical dynamic phantom. Experiment with a moving physical phantom was then carried out to demonstrate the accuracy of the proposed method and the increase of signal‐to‐noise ratio over three‐dimensional PET. Finally, the 4D PET reconstruction was applied to a patient case.Keywords
This publication has 29 references indexed in Scilit:
- Fusion of respiration-correlated PET and CT scans: correlated lung tumour motion in anatomical and functional scansPhysics in Medicine & Biology, 2005
- Four‐dimensional (4D) PET/CT imaging of the thoraxMedical Physics, 2004
- Mapping of the prostate in endorectal coil‐based MRI/MRSI and CT: A deformable registration and validation studyMedical Physics, 2004
- Volume-preserving nonrigid registration of MR breast images using free-form deformation with an incompressibility constraintIEEE Transactions on Medical Imaging, 2003
- PET-CT image registration in the chest using free-form deformationsIEEE Transactions on Medical Imaging, 2003
- Nonrigid registration using free-form deformations: application to breast MR imagesIEEE Transactions on Medical Imaging, 1999
- Cardiac and respiratory gating of list-mode data on a high sensitivity PET scanner, the ECAT EXACT3DJournal of Nuclear Cardiology, 1999
- Real-time system for respiratory-cardiac gating in positron tomographyIEEE Transactions on Nuclear Science, 1998
- Respiratory compensation in projection imaging using a magnification and displacement modelIEEE Transactions on Medical Imaging, 1996
- Electrocardiographic Gating in Positron Emission Computed TomographyJournal of Computer Assisted Tomography, 1979