OPTICAL PRINCIPLES FOR ESTIMATION OF ENDOTHELIAL CELL DENSITY WITH THE NON‐CONTACT SPECULAR MICROSCOPE
Author:
- 1 October 1979
- journal article
- research article
- Published by Wiley in Acta Ophthalmologica
- Vol. 57 (5) , 860-867
Abstract
In non‐contact specular microscopy of corneal endothelium, endothelial cell density is overestimated due to the angle of observation and the curvature of cornea. On the basis of theoretical considerations, it is concluded that if the angle of observation is kept at a small value the curvature of cornea contributes only insignificantly to this overestimation, the major determinant being the angle of observation. When an angle of 46° is chosen between slit illumination and optical axis of the microscope it is calculated that estimates of endothelial cell should be multiplied with a factor of 0.959 to correct for the angle effect. Six eyes were photographed with both a contact and non‐contact specular microscope and endothelial cell density estimated. Mean observed difference in cell counts was +5.5% without correction and +1.2% with correction for angle effect. 95% confidence limits for the difference with correction were ‐3.7 and +6.0% respectively showing that estimates of endothelial cell density obtained with the non‐contact specular microscope agree closely with those obtained by contact specular microscopy when corrected for angle of observation.Keywords
This publication has 5 references indexed in Scilit:
- THE PRECISION OF UNBIASED ESTIMATES OF NUMERICAL DENSITY OF ENDOTHELIAL CELLS IN DONOR CORNEASActa Ophthalmologica, 1978
- Specular Microscopy of Human Corneal Endothelium in VIVOAmerican Journal of Ophthalmology, 1976
- In Vivo Photomicrography of the Corneal EndotheliumArchives of Ophthalmology (1950), 1975
- Cellular membrane activity in the corneal endothelium of the intact eyeCellular and Molecular Life Sciences, 1968
- Die Sichtbarkeit des lebenden HornhautendothelsGraefe's Archive for Clinical and Experimental Ophthalmology, 1920