AC coupled three op-amp biopotential amplifier with active DC suppression
Authors:
- M.A. MayoskyMMM.A. Mayosky
- E.M. SpinelliESE.M. Spinelli
- 1 January 2000
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Biomedical Engineering
- Vol. 47 (12) , 1616-1619
Abstract
A three op-amps instrumentation amplifier (I.A) with active dc suppression is presented. dc suppression is achieved by means of a controlled floating source at the input stage, to compensate electrode and op-amps offset voltages. This isolated floating source is built around an optical-isolated device using a general-purpose optocoupler, working as a photovoltaic generator. The proposed circuit has many interesting characteristics regarding simplicity and cost, while preserving common mode rejection ratio (CMRR) and high input impedance characteristics of the classic three op-amps I.A. As an example, a biopotential amplifier with a gain of 80 dB, a lower cutoff frequency of 0.1 Hz, and a dc input range of +/- 8 mV was built and tested. Using general-purpose op-amps, a CMRR of 105 was achieved without trimmings.Keywords
This publication has 4 references indexed in Scilit:
- Amplifiers for bioelectric events: A design with a minimal number of partsMedical & Biological Engineering & Computing, 1994
- Common mode rejection ratio in differential amplifiersIEEE Transactions on Instrumentation and Measurement, 1991
- An improved buffer for bioelectric signalsIEEE Transactions on Biomedical Engineering, 1989
- A Low-Cost Multichannel Preamplifier for Physiological SignalsIEEE Transactions on Biomedical Engineering, 1987