Abstract
OBJECTIVE: The neurology and physiatry community need improved tools for the evaluation of skeletal muscle condition. Here, we evaluate needle electrical impedance myography (EIM), a new minimally invasive approach to determine muscle status that could ultimately become a bedside tool for the assessment of neuromuscular disorders. APPROACH: We design and study the recording characteristics of tetrapolar EIM needle electrodes combining theory and finite element model simulations. We then use these results to build and pilot in-vivo an EIM needle electrode in the rat gastrocnemius muscle (n = 2). The dielectric properties of muscle are reported (mean ± standard deviation). RESULTS: The numerical simulations show the contribution of subcutaneous fat and muscle tissues to needle EIM data is 97%, respectively; and the sensed volume is 0.70 cm3. Apparent resistivity 277 ± 30 Ωcm and relative permittivity (1 ± 0.4) 105 (dimensionless) measured at 10 kHz are in good agreement with in-vivo dielectric properties reported in the literature. SIGNIFICANCE: The results presented show the feasibility of measuring muscle impedivity in-vivo using a needle electrode from 10 kHz to 1 MHz. The development of needle EIM technology can open up a new field of study in electrodiagnostic medicine, with potential applications to both disease diagnosis and biomarker assessment of therapy.

This publication has 37 references indexed in Scilit: