Low-Dose Quantitative Myocardial Blood Flow Imaging Using 15O-Water and PET Without Attenuation Correction
Authors:
- Mark LubberinkMLMark Lubberink
- Hendrik J. HarmsHHHendrik J. Harms
- Rick HalbmeijerRHRick Halbmeijer
- Stefan de HaanSHStefan de Haan
- Paul KnaapenPKPaul Knaapen
- Adriaan A. LammertsmaALAdriaan A. Lammertsma
- 17 March 2010
- journal article
- Published by Society of Nuclear Medicine in Journal of Nuclear Medicine
- Vol. 51 (4) , 575-580
Abstract
Misalignment between PET and low-dose CT (LD-CT) can cause severe artifacts in cardiac PET/CT because of attenuation-correction errors, even when using slow or cine LD-CT. Myocardial blood flow (MBF), as measured by 15O-water, can be determined from the rate of 15O-water washout from myocardial tissue, which is independent of tissue attenuation. The purpose of the present study was to assess the accuracy of these MBF measurements in the absence of attenuation correction. Methods: Twenty-five patients referred for evaluation of myocardial perfusion underwent 6-min rest and adenosine stress PET scans after the administration of 370 MBq of 15O-water; both scans were followed by slow LD-CT. Data were acquired on a PET/CT scanner and reconstructed by a 3-dimensional row-action maximum likelihood algorithm both with (CTAC) and without (NAC) attenuation correction. An ascending aorta volume of interest was used as input function. MBF and coronary flow reserve (CFR) were calculated for 17 myocardial segments using nonlinear regression of the standard single-tissue-compartment model with corrections for left and right ventricular spillover and perfusable tissue fraction. Results: High correlation (r2 = 0.99 and 0.97, with slopes of 0.96 and 0.91 for rest and stress, respectively) and excellent agreement (intraclass correlation coefficient [ICC], 1.00 and 0.98) between NAC- and CTAC-based MBF values were found. Absolute rest and stress MBF values were 3% and 8%, respectively, lower for NAC scans. The correlation coefficient between all NAC and CTAC CFR values was 0.95 (ICC, 0.95; slope, 0.92) and 0.97 (ICC, 0.99; slope, 1.01) when only CFR values below 2 were considered. Deviations between CTAC and NAC values were smallest for basal segments and increased toward the apex. Conclusion: MBF and CFR can be measured accurately using 15O-water and PET without correcting for attenuation, reducing the effective dose to the patient to 0.8 mSv for a complete rest–stress protocol. This dose is an order of magnitude lower than typical values for 82Rb, 99mTc-methoxyisobutylisonitrile, or CT perfusion scans.Keywords
This publication has 22 references indexed in Scilit:
- Cardiac PET-CT: advanced hybrid imaging for the detection of coronary artery diseaseNetherlands Heart Journal, 2010
- Reproducibility and Accuracy of Quantitative Myocardial Blood Flow Assessment with 82Rb PET: Comparison with 13N-Ammonia PETJournal of Nuclear Medicine, 2009
- Coronary microvascular resistance: methods for its quantification in humansBasic Research in Cardiology, 2009
- Low radiation dose imaging of myocardial perfusion and coronary angiography with a hybrid PET/CT scannerClinical Physiology and Functional Imaging, 2008
- Radiation risk from coronary artery disease imaging: how do different diagnostic tests compare?Heart, 2008
- Cine CT for Attenuation Correction in Cardiac PET/CTJournal of Nuclear Medicine, 2007
- Clinical Myocardial Perfusion PET/CTJournal of Nuclear Medicine, 2007
- New, faster, image-based scatter correction for 3D PETIEEE Transactions on Nuclear Science, 2000
- Noninvasive quantitation of myocardial blood flow in human subjects with oxygen-15-labeled water and positron emission tomographyJACC, 1989
- Quantification of regional myocardial blood flow in vivo with H215O.Circulation, 1984