A hand-held drilling tool for orthopedic surgery
Authors:
- B. AllottaB. Allotta
- G. GiacaloneGGG. Giacalone
- L. RinaldiLRL. Rinaldi
- 1 December 1997
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE/ASME Transactions on Mechatronics
- Vol. 2 (4) , 218-229
Abstract
A novel hand-held drilling tool devoted to orthopedic surgery is presented in this paper. The starting point of the study is the conjecture that the invasiveness of interventions might be reduced by adding sensing, reasoning, and control capabilities to existing tools, in order to obtain controlled penetration in the patient's body and automatic discrimination among layers of different tissues. Due to the particular environment in which the tool is to work, the requirements in terms of human friendliness and safety impose a careful design of the human-machine interface. The proposed approach to the development of a mechatronic tool for surgery is discussed, with emphasis on the functionality and performance of the device and the limited-usage expertise required. New applications for the proposed concepts in nonsurgical environments requiring hand-held tools are foreseen.Keywords
This publication has 9 references indexed in Scilit:
- Robot-assisted invasive orthopaedic surgeryMechatronics, 1996
- A mechatronic drilling tool for ear surgery: A case study of some design characteristicsMechatronics, 1996
- Study on a mechatronic tool for drilling in the osteosynthesis of long bones: Tool/bone interaction, modeling and experimentsMechatronics, 1996
- Manual of INTERNAL FIXATION1991
- Surgical Drilling: Design and Performance of an Improved DrillJournal of Biomechanical Engineering, 1982
- Orthogonal Machining of BoneJournal of Biomechanical Engineering, 1978
- Drilling of boneJournal of Biomechanics, 1976
- A study of the bone machining process—DrillingJournal of Biomechanics, 1976
- Fuzzy setsInformation and Control, 1965