Abstract
We compared Modelflow (MF) estimates of cardiac stroke volume (SV) from the finger pressure‐pulse waveform (Finometer®) with pulsed Doppler ultrasound (DU) of the ascending aorta during acute changes in total peripheral resistance (TPR) in the supine and head‐up‐tilt (HUT) postures. Twenty‐four women were tested during intravenous infusion of 0.005 or 0.01 μg kg−1min−1isoprenaline, 10 or 50 ng kg−1min−1noradrenaline and 0.3 mg sublingual nitroglycerine. Responses to static hand‐grip exercise (SHG), graded lower body negative pressure (LBNP, from −20 to −45 mmHg) and 45 deg HUT were evaluated on separate days. Bland–Altman analysis indicated that SVMFyielded lower estimates than SVDUduring infusion of 0.01 μg kg−1min−1isoprenaline (SVMF92.7 ± 15.5versusSVDU104.3 ± 22.9 ml,P= 0.03) and SHG (SVMF78.8 ± 12.0versusSVDU106.1 ± 28.5 ml,P< 0.01), while larger estimates were recorded with SVMFduring −45 mmHg LBNP (SVMF52.6 ± 10.7versusSVDU46.2 ± 14.5 ml,P= 0.04) and HUT (SVMF59.3 ± 13.6versusSVDU45.2 ± 11.3 ml,P< 0.01). Linear regression analysis revealed a relationship (r2= 0.41,P< 0.01) between the change in TPR from baseline and the between‐methods discrepancy in SV measurements. This relationship held up under all of the experimental protocols (regression for fixed effects,P= 0.46). These results revealed a discrepancy in MF estimates of SV, in comparison with those measured by DU, during acute changes in TPR.