The role of reference body selection in calculating posterior corneal elevation and prediction of keratoconus using rotating Scheimpflug camera
Authors:
- Illés KovácsIKIllés Kovács
- IllésILIllés
- Kata MiháltzKMKata Miháltz
- Mónika EcsedyMEMónika Ecsedy
- János NémethJNJános Németh
- JánosJÁJános
- Zoltán Z NagyZNZoltán Z Nagy
- 25 November 2010
- journal article
- research article
- Published by Wiley in Acta Ophthalmologica
- Vol. 89 (3) , e251-e256
Abstract
: Purpose: To explore the effects of corneal astigmatism and asphericity on posterior elevation values calculated by different reference bodies and to compare their predictive accuracy in the diagnosis of keratoconus.Methods: In this prospective observational case series, 44 eyes of 24 patients with keratoconus and 70 eyes of 41 refractive surgery candidates as a control group were measured by the Pentacam Scheimpflug camera. Discriminating ability and predictors of posterior elevation measurements obtained by best fit toric ellipsoid (BFTE) and best fit sphere (BFS) reference surfaces were compared by receiver operator characteristic curves (ROC) and generalized estimating equation (GEE) models. Bland–Altman plots were used to determine the agreement between different reference surfaces.Results: Receiver operator characteristic curve analysis showed that posterior elevation measured by a BFTE auto had a significantly higher area under ROC curves (0.99) value than BFTE 8‐mm or BFS reference surfaces. ROC analysis identified cut‐off values for BFTE auto (9.5‐μm), for BFTE 8‐mm (10.5‐μm), for BFS auto (16.5‐μm) and for BFS 8‐mm (15.5‐μm) reference surfaces. According to GEE models, corneal cylinder and posterior asphericity had the least effect in toric ellipsoid models. Bland–Altman plots showed a systematic bias at higher values of average posterior elevation measured BFS reference surfaces.Conclusions: Posterior corneal elevation value measured by the Pentacam camera can effectively discriminate keratoconus from normal corneas although measured values and cut‐off points depend on the selection of reference body and corneal asphericity. Toric ellipsoid reference surface seems to be the most sensitive method to differentiate keratoconus.Keywords
This publication has 31 references indexed in Scilit:
- Posterior Elevation in KeratoconusOphthalmology, 2009
- Comparative evaluation of refractive surgery candidates with Placido topography, Orbscan II, Pentacam, and wavefront analysisJournal of Cataract & Refractive Surgery, 2008
- Validity of the keratometric index: Evaluation by the Pentacam rotating Scheimpflug cameraJournal of Cataract & Refractive Surgery, 2008
- Evaluation of anterior segment parameters in keratoconic eyes measured with the Pentacam systemJournal of Cataract & Refractive Surgery, 2007
- Measuring the cornea: the latest developments in corneal topographyCurrent Opinion in Ophthalmology, 2007
- Recent advances in ophthalmic anterior segment imaging: a new era for ophthalmic diagnosis?British Journal of Ophthalmology, 2007
- Corneal-thickness spatial profile and corneal-volume distribution: Tomographic indices to detect keratoconusJournal of Cataract & Refractive Surgery, 2006
- Comparison of Pentacam and Orbscan IIz on Posterior Curvature Topography Measurements in Keratoconus EyesOphthalmology, 2006
- Role of Orbscan II in screening keratoconus suspects before refractive corneal surgeryOphthalmology, 2002
- Bootstrap Methods: Another Look at the JackknifeThe Annals of Statistics, 1979