Sample Processing and Stability for Urine Biomarker Studies
- 28 August 2021
- journal article
- research article
- Published by Oxford University Press (OUP) in The Journal of Applied Laboratory Medicine
- Vol. 6 (6) , 1628-1634
Abstract
Current methods of processing and storing urine samples have not been compared systematically to determine optimal conditions for advancing research on urinary biomarkers. Often, preanalytical handling is nonideal, especially considering the COVID-19 pandemic; consequently, we compared the effects of different short-term storage and processing methods on urinary biomarker measurements. Spot urine samples were collected via a Foley catheter from 20 hospitalized patients from the Yale New Haven Hospital within 48 hours postcardiac surgery. The effects of 3 urine storage and processing methods on biomarkers were tested: (a) 48-hour temporary storage at 4 °C prior to freezing at −80 °C, (b) 48-hour temporary storage at 25 °C prior to freezing at −80 °C, and (c) no centrifugation and immediate storage at −80 °C. Established Meso-Scale Device assay methods were used to measure the urine concentrations of 18 biomarkers: interferon gamma (IFN-ɣ), interleukin (IL)-10, IL-12p70, IL-13, IL-1β, IL-2, IL-4, IL-6, IL-8, IL-18, tumor necrosis factor alpha (TNF-α), epidermal growth factor (EGF), neutrophil gelatinase-associated lipocalin (NGAL), osteopontin (OPN), uromodulin (UMOD), kidney injury molecule-1 (KIM-1), monocyte chemoattractant protein-1 (MCP-1), and chitinase-3-like protein 1 (YKL-40). Measurements of most biomarkers investigated remained stable after temporary storage at 4 °C. IL-6, IL-8, KIM-1, MCP-1, YKL-40, EGF, and NGAL were stable across all 3 processing conditions. IL-12p70 and IL-4 demonstrated significant differences in all tested conditions compared to the reference standard. We identified several notable biomarkers that are robust to variations in preanalytical techniques and can be reliably investigated with nonideal handling conditions.Keywords
Funding Information
- National Institutes of Health (R01HL085757, U01DK082185, UH3DK114866, P30DK079310)
This publication has 9 references indexed in Scilit:
- Recommendations on Acute Kidney Injury Biomarkers From the Acute Disease Quality Initiative Consensus ConferenceJAMA Network Open, 2020
- Tertiary and quaternary structural basis of oxygen affinity in human hemoglobin as revealed by multiscale simulationsScientific Reports, 2017
- Effect of Long-term Storage of Urine Samples on Measurement of Kidney Injury Molecule 1 (KIM-1) and Neutrophil Gelatinase-Associated Lipocalin (NGAL)American Journal of Kidney Diseases, 2014
- Urine Stability Studies for Novel Biomarkers of Acute Kidney InjuryAmerican Journal of Kidney Diseases, 2014
- Determination of uromodulin in human urine: influence of storage and processingNephrology Dialysis Transplantation, 2013
- Evaluation of the ARCHITECT urine NGAL assay: Assay performance, specimen handling requirements and biological variabilityClinical Biochemistry, 2010
- American Society of Nephrology Renal Research ReportJournal of the American Society of Nephrology, 2005
- ROBUST ESTIMATORS OF THE CONCORDANCE CORRELATION COEFFICIENTJournal of Biopharmaceutical Statistics, 2001
- A Concordance Correlation Coefficient to Evaluate Reproducibility1989