MR Imaging and Cardiac Pacemakers: In Vitro Evaluation and in Vivo Studies in 51 Patients at 0.5 T

Authors:
  • CVChristian Vahlhaus
    Christian Vahlhaus
  • GLGerhard Lauck
    Gerhard Lauck
  • ASAlexander V. Smekal
    Alexander V. Smekal
  • MRMarcus Reinke
    Marcus Reinke
  • UHUlrich Hofer
    Ulrich Hofer
  • WBWolfgang Block
    Wolfgang Block
  • CSChristian Schneider
    Christian Schneider
  • JGJürgen Gieseke
    Jürgen Gieseke
  • WJWerner Jung
    Werner Jung
  • HSHans Schild
    Hans Schild

Abstract
PURPOSE: To evaluate the safety and feasibility of magnetic resonance (MR) imaging at 0.5 T in patients with implanted cardiac pacemakers. MATERIALS AND METHODS: Twenty-one models of pacemakers and 44 pacemaker electrodes were exposed to in vitro MR imaging with continuous registration of pacemaker output and temperature at the lead tip. Prior to MR imaging examination, pacemakers were programmed to an asynchronous mode (A00, V00, or D00). Pacemakers were examined before and after MR imaging. Forty-four patients with implanted pacemakers underwent 51 MR imaging examinations under cardiologic surveillance, continuous electrocardiography, pulse oximetry, and capnographic monitoring. RESULTS: MR imaging was safely performed in all patients. None of the pacemakers displayed a pacing dysfunction at MR imaging. No changes occurred in the programmed parameters in any device tested in vivo or in vitro. Maximum increases in the temperature at the lead tips were 8.90°C at a specific absorption rate (SAR) of 0.6 W/k...