An Automated Microfluidic System for Morphological Measurement and Size-Based Sorting of C. Elegans
Authors:
- Xianke DongXDXianke Dong
- Pengfei SongPSPengfei Song
- Xinyu LiuXinyu Liu
- 8 March 2019
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Nanobioscience
- Vol. 18 (3) , 373-380
Abstract
This paper reports a vision-based automated microfluidic system for morphological measurement and size-based sorting of the nematode worm C. elegans. Exceeding the capabilities of conventional worm sorting microfluidic devices purely relying on passive sorting mechanisms, our system is capable of accurate measurement of the worm length/width and active sorting of worms with the desired sizes from a mixture of worms with different body sizes. This function is realized based on the combination of real-time, vision-based worm detection and sizing algorithms and automated on-chip worm manipulation. A double-layer microfluidic device with computer-controlled pneumatic valves is developed for sequential loading, trapping, vision-based sizing, and sorting of single worms. To keep the system operation robust, vision-based algorithms on detecting multi-worm loading and worm sizing failure have also been developed. We conducted sorting experiments on 319 worms and achieved an average sorting speed of 10.4 worms per minute (5.8 s/worm) with an operation success rate of 90.3%. This system will facilitate the worm biology studies where body size measurement and size-based sorting of many worms are needed.Keywords
Funding Information
- Natural Sciences and Engineering Research Council of Canada (RGPIN-2017-06374, RGPAS-2017-507980, DGDND-2017-00001)
- Chinese Scholarship Council
This publication has 33 references indexed in Scilit:
- Proprioceptive Coupling within Motor Neurons Drives C. elegans Forward LocomotionNeuron, 2012
- Electrical sorting of Caenorhabditis elegansLab on a Chip, 2012
- Running Worms: C. elegans Self-Sorting by ElectrotaxisPLOS ONE, 2011
- Non-stringent tissue-source requirements for BMP ligand expression in regulation of body size inCaenorhabditis elegansGenetics Research, 2011
- Microfluidics for the analysis of behavior, nerve regeneration, and neural cell biology in C. elegansCurrent Opinion in Neurobiology, 2009
- crm-1 facilitates BMP signaling to control body size in Caenorhabditis elegansDevelopmental Biology, 2007
- Microfluidic system for on-chip high-throughput whole-animal sorting and screening at subcellular resolutionProceedings of the National Academy of Sciences, 2007
- Regulation of Caenorhabditis elegans body size and male tail development by the novel gene lon-8BMC Developmental Biology, 2007
- A Caenorhabditis elegans Wild Type Defies the Temperature–Size Rule Owing to a Single Nucleotide Polymorphism in tra-3PLOS Genetics, 2007
- Regulation of Body Size and Behavioral State of C. elegans by Sensory Perception and the EGL-4 cGMP-Dependent Protein KinaseNeuron, 2002