Multivariate Calibration and Instrument Standardization for the Rapid Detection of Diethylene Glycol in Glycerin by Raman Spectroscopy
- 1 March 2011
- journal article
- research article
- Published by SAGE Publications in Applied Spectroscopy
- Vol. 65 (3) , 334-341
Abstract
The transfer of a multivariate calibration model for quantitative determination of diethylene glycol (DEG) contaminant in pharmaceutical-grade glycerin between five portable Raman spectrometers was accomplished using piecewise direct standardization (PDS). The calibration set was developed using a multi-range ternary mixture design with successively reduced impurity concentration ranges. It was found that optimal selection of calibration transfer standards using the Kennard–Stone algorithm also required application of the algorithm to multiple successively reduced impurity concentration ranges. Partial least squares (PLS) calibration models were developed using the calibration set measured independently on each of the five spectrometers. The performance of the models was evaluated based on the root mean square error of prediction (RMSEP), calculated using independent validation samples. An F-test showed that no statistical differences in the variances were observed between models developed on different instruments. Direct cross-instrument prediction without standardization was performed between a single primary instrument and each of the four secondary instruments to evaluate the robustness of the primary instrument calibration model. Significant increases in the RMSEP values for the secondary instruments were observed due to instrument variability. Application of piecewise direct standardization using the optimal calibration transfer subset resulted in the lowest values of RMSEP for the secondary instruments. Using the optimal calibration transfer subset, an optimized calibration model was developed using a subset of the original calibration set, resulting in a DEG detection limit of 0.32% across all five instruments.Keywords
This publication has 17 references indexed in Scilit:
- Transfer of Multivariate Calibration Models2009
- Oversulfated chondroitin sulfate is a contaminant in heparin associated with adverse clinical eventsNature Biotechnology, 2008
- A comparative study of calibration transfer methods for determination of gasoline quality parameters in three different near infrared spectrometersAnalytica Chimica Acta, 2008
- Raman spectroscopy of glycerol/D2O solutionsVibrational Spectroscopy, 2007
- Robust modeling for multivariate calibration transfer by the successive projections algorithmChemometrics and Intelligent Laboratory Systems, 2005
- Transfer of multivariate calibration models: a reviewChemometrics and Intelligent Laboratory Systems, 2002
- Standardisation and Calibration Transfer for near Infrared Instruments: A ReviewJournal of Near Infrared Spectroscopy, 2001
- Cryogenic Raman spectroscopy of glycerolJournal of Raman Spectroscopy, 2000
- Standardisation of near-infrared spectrometric instruments: A reviewVibrational Spectroscopy, 1996
- Multivariate calibration standardizationChemometrics and Intelligent Laboratory Systems, 1994