A solid state microdosimeter based on a monolithic silicon telescope
Authors:
- P. ColauttiPCP. Colautti
- A. FazziAFA. Fazzi
- D. MoroDMD. Moro
- A. PolaAPA. Pola
- 1 December 2006
- journal article
- research article
- Published by Oxford University Press (OUP) in Radiation Protection Dosimetry
- Vol. 122 (1-4) , 382-386
Abstract
A monolithic silicon telescope, consisting of a ΔE and an E stage-detector (∼1.9 μm and 500 μm thick, respectively), was coupled to a polyethylene converter in order to investigate the feasibility of a solid state microdosimeter with respect to the field-funnelling effect. This work discusses the preliminary results of an analytical approach for the correction of a spectrum measured with this silicon-based microdosimeter for tissue-equivalence and geometrical effects. The device was irradiated with 2.7 MeV monoenergetic neutrons at the INFN-Laboratori Nazionali di Legnaro (Legnaro, Italy). The non tissue-equivalence of silicon was corrected by exploiting the signals generated in the E-stage. The correction for the sensitive volume geometry was optimised by taking into account the track length distribution of the recoil-protons generated in the converter. The derived dose distribution of the energy imparted per event was compared to the one measured with a cylindrical tissue-equivalent proportional counter (TEPC). The agreement is satisfactory.Keywords
This publication has 8 references indexed in Scilit:
- Secondary photon fields produced in accelerator-based sources for neutron generationRadiation Protection Dosimetry, 2005
- A feasibility study of a solid-state microdosimeterApplied Radiation and Isotopes, 2005
- Solid state microdosimetryNuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms, 2001
- Electron-Photon Transport in FLUKA: Status2001
- A new large area monolithic silicon telescopeNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1999
- Criteria for the Equivalence of Spherical and Cylindrical Proportional Counters in MicrodosimetryRadiation Research, 1981
- A field-funneling effect on the collection of alpha-particle-generated carriers in silicon devicesIEEE Electron Device Letters, 1981
- Concepts of microdosimetryRadiation and Environmental Biophysics, 1975