An effective AI model for automatically detecting arteriovenous fistula stenosis
Open Access
- 17 October 2023
- journal article
- research article
- Published by Springer Nature in Scientific Reports
- Vol. 13 (1) , 1-11
Abstract
In this study, a novel artificial intelligence (AI) model is proposed to detect stenosis in arteriovenous fistulas (AVFs) using inexpensive and non-invasive audio recordings. The proposed model is a combination of two new input features based on short-time Fourier transform (STFT) and sample entropy, as well as two associated classification models (ResNet50 and ANN). The model’s hyper-parameters were optimized through the use of the design of the experiment (DOE). The proposed AI model demonstrates high performance with all essential metrics, including sensitivity, specificity, accuracy, precision, and F1-score, exceeding 0.90 at detecting stenosis greater than 50%. These promising results suggest that our approach can lead to new insights and knowledge in this field. Moreover, the robust performance of our model, combined with the affordability of the audio recording device, makes it a valuable tool for detecting AVF stenosis in home-care settings.Keywords
This publication has 30 references indexed in Scilit:
- Novel Noninvasive Approach for Detecting Arteriovenous Fistula StenosisIEEE Transactions on Biomedical Engineering, 2014
- Displaying statistical point estimates using the leading-digit ruleIIE Transactions, 2011
- Omitting Meaningless Digits in Point Estimates: The Probability Guarantee of Leading-Digit RulesOperations Research, 2009
- The Arteriovenous FistulaJournal of the American Society of Nephrology, 2003
- Increasing arteriovenous fistulas in hemodialysis patients: Problems and solutionsKidney International, 2002
- The empirical mode decomposition and the Hilbert spectrum for nonlinear and non-stationary time series analysisProceedings of the Royal Society A: Mathematical, Physical and Engineering Sciences, 1998
- Noninvasive acoustical detection of coronary artery disease: a comparative study of signal processing methodsIEEE Transactions on Biomedical Engineering, 1993
- Universal Approximation Using Radial-Basis-Function NetworksNeural Computation, 1991
- Some comments on maximum entropy spectral analysis of time seriesComputers in Physics, 1990
- Noninvasive detection of intracranial vascular lesions using an electronic stethoscopeJournal of Neurosurgery, 1984