Improved wear performance with crosslinked UHMWPE and zirconia implants in knee simulation
Authors:
Open Access
- 1 January 2006
- journal article
- research article
- Published by Medical Journals Sweden AB in Acta Orthopaedica
- Vol. 77 (3) , 505-511
Abstract
Background Suggestions for improved wear performance of total knee replacements have included replacement of standard CoCr femoral components with ceramic and replacement of 3.5-Mrad ultra-high molecular weight polyethylene (UHMWPE) inserts with 5- or 7-Mrad UHMWPE inserts. The ceramic materials used clinically have included alumina, zirconia ceramic and oxidized zirconium. Patients and method We compared both CoCr and zirconia versions of the Bi-Surface knee replacement in a 6-station knee simulator using alpha calf serum for lubrication (20 mg protein per mL) to evaluate the relative bearing performance. Results We studied the 4-way knee simulation of implant materials: zirconia ceramic, CoCr, 3.5-Mrad UHMWPE, and 7-Mrad UHMWPE. With CoCr femoral components, the 7-Mrad UHMWPE resulted in a 5- to 8-fold reduction in wear compared to the 3.5-Mrad insert. With the 3.5-Mrad insert, the zirconia bearing provided approximately 4-fold wear reduction compared to CoCr. These wear rates with standard UHMWPE were similar to published wear studies on entire knees. With the exception of the CoCr/7-Mrad and ZrO2/3.5-Mrad combinations, the wear differences were statistically significant. Interpretation The ZrO2/7-Mrad UHMWPE combination gave the best performance, with no measurable wear over the 5.5 million cycle test duration.Keywords
This publication has 17 references indexed in Scilit:
- Comparison of the Wear Properties of Polyethylene Plate in Total Knee Prostheses (TKP) using Different Femoral Component MaterialsKey Engineering Materials, 2003
- Catastrophic failure of the Elite Plus total hip replacement, with a Hylamer acetabulum and Zirconia ceramic femoral headThe Journal of Bone and Joint Surgery. British volume, 2002
- Phase transformation of a zirconia ceramic head after total hip arthroplastyThe Journal of Bone and Joint Surgery. British volume, 2001
- A comparison of polyethylene wear in hips with cobalt-chrome or zirconia headsThe Journal of Bone and Joint Surgery. British volume, 2001
- Correlation of wear debris-induced osteolysis and revision with volumetric wear-rates of polyethylene: A survey of 8 reports in the literatureActa Orthopaedica, 2001
- CERAMICS IN ORTHOPAEDICSThe Journal of Bone and Joint Surgery. British volume, 2000
- Zirconia as a Sliding MaterialClinical Orthopaedics and Related Research, 2000
- Ageing of Zirconia: Everything You Always Wanted to KnowKey Engineering Materials, 2000
- Hip-simulator ranking of polyethylene wear:Comparisons between ceramic heads of different sizesActa Orthopaedica, 1996
- Long-Term Results of the Total Condylar Knee ArthroplastyClinical Orthopaedics and Related Research, 1993