Effect of Application of a Bio-Adhesive on the Removal Torque Value and Rotational Misfit at the Implant–Abutment Junction: An In Vitro Study
Open Access
- 12 November 2021
- Vol. 14 (22) , 6832
Abstract
The aim of this study was to assess the effect of application of a recently developed bio-adhesive (Impladhesive) to abutment screw threads on the removal torque value and rotational misfit at the implant–abutment junction. This in vitro study evaluated 20 implant fixtures and 20 straight abutments. Specimens were randomly divided into two groups (n = 10) with/without adhesive application. In the adhesive group, the abutment was dipped in Impladhesive before torquing. In the control group, the abutment was torqued conventionally without adhesive application. The removal torque value was recorded after completion of the cyclic loading of 500,000 cycles with 2 Hz frequency and 75 N load. Rotational misfit was recorded using a video measuring machine. After applying the torque, the change in the bisector angle on the abutment hex was recorded for each implant. The biocompatibility of Impladhesive was evaluated using a MTT cell vitality assay. Normal distribution of data was assessed using the Kolmogorov–Smirnov test. Data were analyzed using a t-test and Pearson’s correlation coefficient The application of Impladhesive at the implant–abutment interface resulted in significantly greater mean removal torque value compared to the control group (p = 0.008). In addition, the mean rotational misfit at the implant–abutment interface was significantly lower in the use of Impladhesive compared to the control group (p = 0.001). In addition, the cell vitality was found to be greater than 80% at all evaluated time points. It can be concluded that the application of Impladhesive on the abutment screw significantly decreased rotational misfit and increased the removal torque value. Future studies are needed to evaluate the efficacy of this bio-adhesive an in vivo setting.Keywords
This publication has 31 references indexed in Scilit:
- Correlation between microleakage and screw loosening at implant-abutment connectionThe Journal of Advanced Prosthodontics, 2014
- Bacterial Colonization of the Implant–Abutment Interface Using an In Vitro Dynamic Loading ModelJournal of Periodontology, 2011
- Cell Sensitivity Assays: The MTT Assay2011
- Removal torque of zirconia abutment screws under dry and wet conditionsBrazilian Dental Journal, 2010
- Probabilistic analysis of preload in the abutment screw of a dental implant complexThe Journal of Prosthetic Dentistry, 2008
- The Implant‐Abutment Interface of Alumina and Zirconia AbutmentsClinical Implant Dentistry and Related Research, 2008
- A systematic review of the 5‐year survival and complication rates of implant‐supported single crownsClinical Oral Implants Research, 2007
- Marginal and internal fit of all‐ceramic CAD/CAM crown‐copings on chamfer preparationsJournal of Oral Rehabilitation, 2005
- Preventing loosening of implant abutment screwsThe Journal of Prosthetic Dentistry, 1996
- EVALUATION OF THREE SLIP FIT HEXAGONAL IMPLANTSImplant Dentistry, 1996