Contact force and impedance decrease during ablation depends on catheter location and orientation: insights from pulmonary vein isolation using a contact force-sensing catheter
- 30 April 2015
- journal article
- clinical trial
- Published by Springer Nature in Journal of Interventional Cardiac Electrophysiology
- Vol. 43 (3) , 297-306
Abstract
Contact force (CF) sensing during radiofrequency (RF) ablation allows controlling lesion size. The aim of this study was to analyze the impact of catheter tip location and orientation on the association of CF and impedance decrease. We retrospectively analyzed RF applications from 32 patients undergoing catheter ablation for paroxysmal atrial fibrillation using a force-sensing catheter and 3D mapping system. CF, catheter location and orientation relative to the tissue during ablation as well as the absolute impedance decrease during the first 20 s of ablation as a surrogate for lesion effectiveness were analyzed for 791 RF applications. While a higher CF was achieved around the right pulmonary veins (12.5 vs. 11.4 g, p = 0.045), a lower median absolute impedance decrease within the first 20 s was seen around the right veins compared to the left veins (9.3 vs. 10.2 Ω, p = 0.02). With different catheter orientations relative to the tissue, higher CF and impedance decrease was seen when the catheter was orientated parallel or oblique to the tissue (30°–145°) as compared perpendicularly (0–30°) with a median CF of 13.2 vs. 8.0 g (p < 0.001) and a median impedance decrease during the first 20 s of 11 vs. 7 Ω (p < 0.001). Importantly, achieved CF, baseline impedance, catheter orientation and location all independently predicted the initial absolute and relative impedance decrease in a multivariable linear regression model (p < 0.05). The effectiveness of RF ablation lesions, as assessed by the initial impedance decrease, is not only dependent on the achieved catheter CF but also on catheter orientation and location.Keywords
This publication has 22 references indexed in Scilit:
- Contact Geometry Affects Lesion Formation in Radio-Frequency Cardiac Catheter AblationPLOS ONE, 2013
- Contact force sensing technology identifies sites of inadequate contact and reduces acute pulmonary vein reconnection: A prospective case control studyInternational Journal of Cardiology, 2013
- Electrical Reconnection After Pulmonary Vein Isolation Is Contingent on Contact Force During Initial TreatmentCirculation: Arrhythmia and Electrophysiology, 2013
- The relationship between contact force and clinical outcome during radiofrequency catheter ablation of atrial fibrillation in the TOCCATA studyHeart Rhythm, 2012
- Biophysics of Radiofrequency Lesion Formation2011
- Area Under the Real‐Time Contact Force Curve (Force–Time Integral) Predicts Radiofrequency Lesion Size in an In Vitro Contractile ModelJournal of Cardiovascular Electrophysiology, 2010
- Early Assessment of Biophysical Parameters Predicts Lesion Formation During RF Energy Delivery In VitroPacing and Clinical Electrophysiology, 2010
- Local unipolar and bipolar electrogram criteria for evaluating the transmurality of atrial ablation lesions at different catheter orientations relative to the endocardial surfaceHeart Rhythm, 2010
- Endocardial impedance mapping during circumferential pulmonary vein ablation of atrial fibrillation differentiates between atrial and venous tissueHeart Rhythm, 2006
- Lesion Dimensions During Temperature-Controlled Radiofrequency Catheter Ablation of Left Ventricular Porcine MyocardiumCirculation, 1999