Development of an ultrahigh vacuum scanning tunneling microscope cooled by superfluid 4He
Authors:
- 26 June 2001
- journal article
- research article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 72 (7) , 2977-2983
Abstract
We have developed an ultrahigh vacuum (UHV) scanning tunneling microscope (STM) cooled by superfluid This microscope is integrated with a solenoid and split-pair superconducting magnet. The STM can be operated at in a rotatable magnetic field of up to 8 T perpendicular to, and 1.5 T parallel to the sample surface. Moreover, tips and samples may be changed without venting the UHV system. The performance of the STM was confirmed by obtaining atomic resolution images of Si reconstructed surfaces at low temperatures. The STM performance was unaffected by the application of high magnetic fields.
Keywords
This publication has 15 references indexed in Scilit:
- Anomalous flipping motions of buckled dimers on the Si(001) surface at 5 KPhysical Review B, 2000
- Kondo Scattering Observed at a Single Magnetic ImpurityPhysical Review Letters, 1998
- A low-temperature ultrahigh-vacuum scanning tunneling microscope with rotatable magnetic fieldReview of Scientific Instruments, 1997
- Probing the Local Effects of Magnetic Impurities on SuperconductivityScience, 1997
- Near-field imaging of atom diffraction gratings: The atomic Talbot effectPhysical Review A, 1995
- Low-temperature scanning-tunneling-microscopy observations of the Si(001) surface with a low surface-defect densityPhysical Review B, 1994
- Integrated cryogenic scanning tunneling microscopy and sample preparation systemReview of Scientific Instruments, 1994
- Advanced chaos forecastingPhysical Review E, 1994
- Stability of the wurtzite-type structure under high pressure: GaN and InNPhysical Review B, 1994
- Direct observation of an increase in buckled dimers on Si(001) at low temperaturePhysical Review Letters, 1992