Defects in ProTaper S1 instruments after clinical use: longitudinal examination
Authors:
- 11 July 2005
- journal article
- Published by Wiley in International Endodontic Journal
- Vol. 38 (8) , 550-557
Abstract
Aim To evaluate defects in ProTaper shaping instrument S1 after a defined schedule of clinical use.Methodology Among all ProTaper files discarded from an endodontic clinic at a stomatological school in China over a period of 17 months, 122 S1 instruments were collected. They were ultrasonically cleaned, autoclaved and then examined in the laboratory. Any instrument separation was noted; the average length involved was determined. The 0.5 mm region on either side of this length of discarded but intact instruments was examined circumferentially by scanning electron microscope. The region adjacent to the broken end of the fractured instruments was also examined in the same way.Results One specimen was lost during processing. Of the remaining 121 instruments, 27 were separated with a mean fracture length of 3.67 mm from the tip. Of these, two files showed macroscopically torsional fracture and the others showed signs of flexural fatigue. Only one file that had not fractured showed visible unwinding defects. Examination of the 3.1–4.1 mm region of other unfractured instruments revealed the presence of microcracks, surface debris, pitting and/or wear of their cutting edges. Some debris particles seemed to have been trapped in crack‐like structures.Conclusions Multiple use of ProTaper S1 pre‐disposed the instrument to microcrack formation and wear of the cutting edges. There was a low prevalence of plastic deformation and most ProTaper S1 instruments failed without discernible sign of unwinding of the flutes. Further studies should address the mode of failure and the role of debris particles in the fracture mechanism.Keywords
This publication has 24 references indexed in Scilit:
- Proposed Role of Embedded Dentin Chips for the Clinical Failure of Nickel-Titanium Rotary InstrumentsJournal of Endodontics, 2004
- K3 Endo, ProTaper, and ProFile Systems: Breakage and Distortion in Severely Curved Roots of MolarsJournal of Endodontics, 2004
- Evaluation of Single-use Rotary Nickel-titanium InstrumentsJournal of Endodontics, 2003
- Scanning Electron Microscope Observations of New and Used Nickel-Titanium Rotary FilesJournal of Endodontics, 2003
- Comparative Analysis of Torsional and Bending Stresses in Two Mathematical Models of Nickel-Titanium Rotary Instruments: ProTaper versus ProFileJournal of Endodontics, 2003
- Cyclic Fatigue of Nickel-Titanium Rotary Instruments after Clinical Use with Low-and High-Torque Endodontic MotorsJournal of Endodontics, 2001
- Comparison of Nickel-Titanium File Distortion Using Electric and Air-Driven HandpiecesJournal of Endodontics, 2001
- Effect of rotational speed on nickel-titanium file distortionJournal of Endodontics, 1999
- Wear of nickel-titanium lightspeed instruments evaluated by scanning electron microscopyJournal of Endodontics, 1999
- Analysis of forces developed during mechanical preparation of extracted teeth using Profile NiTi rotary instrumentsInternational Endodontic Journal, 1999