Auto-CPAP therapy based on the forced oscillation technique. Auto-CPAP-Therapie auf der Basis der Oszilloresistometrie
Authors:
- J. H. FickerJFJ. H. Ficker
- C. F. ClarenbachCCC. F. Clarenbach
- C. NeukirchnerCNC. Neukirchner
- F. S. FuchsFFF. S. Fuchs
- G. H. WiestGWG. H. Wiest
- S. Pour SchahinSSS. Pour Schahin
- I. A. HarschIHI. A. Harsch
- E. G. HahnEHE. G. Hahn
- 1 January 2003
- journal article
- clinical trial
- Published by Walter de Gruyter GmbH in Biomedical Engineering / Biomedizinische Technik
- Vol. 48 (3) , 68-72
Abstract
Autoadjusting CPAP devices (APAP) are designed to continuously adjust the positive pressure to the required levels, and thus increase treatment quality and patient compliance. The results of APAP treatment strongly depend on the control mechanism of the respective APAP device. In agreement with other working groups, we have recently shown that the forced oscillation technique (FOT) is capable of detecting incipient upper airway obstruction prior to physiological reactions such as the onset of increasing esophageal pressure swings or microarousals. Therefore we studied efficacy and acceptance of a novel APAP device controlled exclusively by FOT. 100 consecutive patients with OSAS confirmed by polysomnography (mean AHI 47.9 +/- 22.6) and daytime sleepiness (Epworth sleepiness scale, ESS 12.6 +/- 3.9) were randomized to either APAP treatment (n = 50) or conventional CPAP treatment (n = 50). Polysomnographies were performed at the second treatment night and subjective sleepiness (modified ESS) was established in the morning. The respiratory disturbance was largely normalized in both treatment groups in the second treatment night (AHI 4.7 +/- 5.3 vs. 3.7 +/- 3.4; n.s.). Both groups showed largely improved sleep profiles and had markedly reduced ESS-scores (6.6 +/- 3.6 vs. 7.0 +/- 3.4; n.s.). The mean treatment pressure during APAP was significantly lower than during CPAP treatment (6.0 +/- 2.0 vs. 9.0 +/- 1.8 mbar; p < 0.001). There were no significant differences between APAP and CPAP treatment in any parameter of efficacy or acceptance. APAP treatment with this device controlled exclusively by FOT is well accepted by the patients and permits an adequate treatment of OSAS without the need for invidiual CPAP titration.Keywords
This publication has 14 references indexed in Scilit:
- Autoadjusting CPAP Therapy Based on Impedance Efficacy, Compliance and AcceptanceAmerican Journal of Respiratory and Critical Care Medicine, 2001
- Self-adjusting Nasal Continuous Positive Airway Pressure Therapy Based on Measurement of ImpedanceChest, 1999
- Assessment of Airflow Obstruction during CPAP by Means of Forced Oscillation in Patients with Sleep ApneaAmerican Journal of Respiratory and Critical Care Medicine, 1998
- A system to generate simultaneous forced oscillation and continuous positive airway pressureEuropean Respiratory Journal, 1997
- Efficacy of Auto-CPAP in the Treatment of Obstructive Sleep Apnea/Hypopnea SyndromeAmerican Journal of Respiratory and Critical Care Medicine, 1996
- Determinants of Effective Continuous Positive Airway Pressure in Obstructive Sleep Apnea. Role of Respiratory EffortAmerican Journal of Respiratory and Critical Care Medicine, 1995
- Effect of continuous positive airway pressure treatment on daytime function in sleep apnoea/hypopnoea syndromeThe Lancet, 1994
- Feasibility of a Self-Setting CPAP MachineSLEEP, 1993
- Relations Between Sleep Stage, Posture and Effective Nasal CPAP Levels in OSASLEEP, 1992
- A Preliminary Report from the Sleep Disorders Atlas Task Force of the American Sleep Disorders AssociationSLEEP, 1992