Effects of bladder shape on accuracy of measurement of bladder volume using portable ultrasound scanner and development of correction method
- 8 January 2019
- journal article
- research article
- Published by Wiley in Neurourology and Urodynamics
- Vol. 38 (2) , 653-659
Abstract
Aims: To compare the accuracy of using a bladder scanner to measure bladder volume through intermittent catheterization (IC) in patients and to introduce the Bladder Deformation Index (BDI) to develop a correction method.Methods: Bladder volume was assessed by a nurse with the scanner. A second nurse catheterized the patient's bladder. A third nurse measured the urine volume in a 500‐mL or 1000‐mL graduated cylinder.Results: Sixty one patients were included and 590 pairs of data were obtained. The mean bladder volume measured using a scanner and IC was (332.3 ± 156.1) mL and (339.1 ± 158.8) mL. The mean absolute difference was 30.8 mL. The correlation coefficient was 0.929. Patients were classified into 2 groups depending on whether they had undergone augmentation cystoplasty. The mean absolute difference was 109.2 and 20.4 mL. The correlation coefficient was 0.712 and 0.981. According to the BDI, bladders can be classified into 3 groups. The mean absolute difference was 21.9, 60.4, and 109.4 mL. The correlation coefficient was 0.970, 0.839, and 0.783. The linear regression equations of Grade I and Grade II were Y = 1.11X + 3.1 and Y = 0.76X + 161.5.Conclusions: The results showed that bladder shape plays a critical role in accuracy which is inversely associated with BDI. This degree of accuracy is sufficient; especially measurement adjusted using the linear regression equation in patients with high BDI. However, although the preliminary results of the study are promising, a large‐scale prospective study should be needed to address the validation of the data in the future.Keywords
Funding Information
- National Natural Science Foundation of China (81570688)
This publication has 16 references indexed in Scilit:
- Interfractional variation in bladder volume and its impact on cervical cancer radiotherapy: Clinical significance of portable bladder scannerMedical Physics, 2016
- Diagnosis of Postoperative Urinary Retention Using a Simplified Ultrasound Bladder MeasurementAnesthesia & Analgesia, 2015
- Is portable three-dimensional ultrasound a valid technique for measurement of postpartum urinary bladder volume?Taiwanese Journal of Obstetrics and Gynecology, 2014
- Bladder filling variations during concurrent chemotherapy and pelvic radiotherapy in rectal cancer patients: early experience of bladder volume assessment using ultrasound scannerRadiation Oncology Journal, 2013
- Accuracy of Bladder Scanning in the Assessment of Postvoid Residual VolumeJournal of Obstetrics and Gynaecology Canada, 2009
- Cystic pelvic pathology presenting as falsely elevated postvoid residual urine measured by portable ultrasound bladder scanning: report of 3 cases and review of the literatureUrology, 2000
- Bladder volume determination: Portable 3-D versus stationary 2-D ultrasound deviceArchives of Physical Medicine and Rehabilitation, 2000
- Three-dimensional ultrasound device for rapid determination of bladder volumeUrology, 1997
- Determining residual urine volumes using a portable ultrasonographic deviceArchives of Physical Medicine and Rehabilitation, 1993
- ULTRASOUND MEASUREMENT OF URINE VOLUME OF CHILDREN WITH NEUROGENIC BLADDERDevelopmental Medicine & Child Neurology, 1990