Randomised, crossover comparison of the single‐use SoftSeal ™ and the LMA Unique ™ laryngeal mask airways
Authors:
Open Access
- 14 March 2005
- journal article
- clinical trial
- Published by Wiley in Anaesthesia
- Vol. 60 (4) , 354-359
Abstract
Summary: We performed a randomised, crossover, single‐blind trial among 168 patients, to compare the single‐use SoftSeal ™ and LMA Unique ™ laryngeal mask airways in spontaneously breathing adults. Size‐3 and ‐4 laryngeal mask airways, inserted by experienced anaesthetists, performed equivalently for successful first‐time placement (148 (91%) vs 155 (96%) for the SoftSeal ™ and LMA Unique ™ , respectively). The SoftSeal ™ was more often rated as difficult to insert (27 (17%) vs 4 (3%); p < 0.001) and was more likely to show evidence of mucosal trauma after the first insertion (14 (10%) vs 5 (4%); OR 1.3 (95% CI 1.3–11.3); p < 0.05). The fibreoptic view of the larynx was better through the SoftSeal ™ (vocal cords not visible in 27 (17%) vs 44 (27%); p < 0.05) and it more frequently provided a ventilation seal at 20 cmH 2 O (93 (59%) vs 62 (39%); OR 2.15 (CI 1.44–3.21); p < 0.001). In contrast to the LMA Unique ™ , its cuff pressure did not increase during nitrous oxide anaesthesia (median (IQR [range]) decrease 3 (− 20–23 [‐40–94]) cmH 2 O vs increase of 16 (−2–39 [−54–112]) cmH 2 O; p < 0.01). Both devices were equivalent for the success of first‐time insertion and performed satisfactorily clinically. There were some performance differences, but either appears suitable for airway management in spontaneously breathing patients.Keywords
This publication has 24 references indexed in Scilit:
- The Laryngeal Mask Airway Unique??? versus the Soft Seal??? Laryngeal Mask: A Randomized, Crossover Study in Paralyzed, Anesthetized PatientsAnesthesia & Analgesia, 2004
- Pharyngeal Mucosal Pressure and PerfusionAnesthesiology, 1999
- Mucosal Pressure, Mechanism of Seal, Airway Sealing Pressure, and Anatomic Position for the Disposable Versus Reusable Laryngeal Mask AirwaysAnesthesia & Analgesia, 1999
- A Prospective Evaluation of Clinical Tests for Placement of Laryngeal Mask AirwaysAnesthesiology, 1998
- A Comparison of the Disposable Versus the Reusable Laryngeal Mask Airway in Paralyzed Adult PatientsAnesthesia & Analgesia, 1998
- Clinical assessment of the single use laryngeal mask airway—the LMA-uniqueBritish Journal of Anaesthesia, 1998
- Insertion methods of the laryngeal mask airwayAnaesthesia, 1996
- Laryngeal mask airway: a more successful method of insertionJournal of Clinical Anesthesia, 1995
- NEUROMUSCULAR BLOCK AND INSERTION OF THE LARYNGEAL MASK AIRWAYBritish Journal of Anaesthesia, 1993
- The effect of nitrous oxide on laryngeal mask cuff pressureAnaesthesia, 1992