Voltammetric Ion-Selective Electrodes in Thin-Layer Samples: Absolute Detection of Ions Using Ultrathin Membranes
Open Access
- 5 January 2024
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 96 (3) , 1147-1155
Abstract
Calibration-free sensors are generally understood as analytical tools with no need for calibration apart from the initial one (i.e., after its fabrication). However, an "ideal" and therefore "more restricted" definition of the concept considers that no calibration is necessary at all, with the sensor being capable of directly providing the analyte concentration in the sample. In the electroanalysis field, investigations have been directed to charge-based readouts (i.e., coulometry) that allow for concentration calculation via the Faraday Law: The sample volume must be precisely defined and the absoluteness of the electrochemical process in which the analyte is involved must be ensured (i.e., the analyte in the sample is ∼100% converted/transported). Herein, we report on the realization of calibration-free coulometric ISEs based on ultrathin ion-selective membranes, which is demonstrated for the detection of potassium ions (K+). In essence, the K+ transfer at the membrane-sample interface is modulated by the oxidation state of the conducting polymer underlying the membrane. The accumulation/release of K+ to/from the membrane is an absolute process owing to the confinement of the sample to a thin-layer domain (thickness of + present in ca. 5 μL sample volume. The system is interrogated with cyclic voltammetry to obtain peaks related to the K+ transfer that can be treated charge-wise. The conceptual and technical innovative steps developed here made the calibration-free detection of K+ possible in artificial and real samples with acceptable accuracy (<10% difference compared with the results obtained from a current-based calibration and ion chromatography). The charge-based analysis does not depend on temperature and appeared to be repetitive, reproducible, and reversible in the concentration range from 1 to 37.5 μM, with an average coulometry efficiency of 96%.Keywords
Funding Information
- European Research Council (851957)
- China Scholarship Council
This publication has 26 references indexed in Scilit:
- Exhaustive Thin-Layer Cyclic Voltammetry for Absolute Multianalyte Halide DetectionAnalytical Chemistry, 2014
- Thin Layer Coulometry Based on Ion-Exchanger Membranes for Heparin Detection in Undiluted Human BloodAnalytical Chemistry, 2014
- A low-cost thin layer coulometric microfluidic device based on an ion-selective membrane for calcium determinationThe Analyst, 2013
- Thin layer coulometric determination of nitrate in fresh watersAnalytica Chimica Acta, 2012
- Thin layer coulometry ion sensing protocol with potassium-selective membrane electrodesElectrochimica Acta, 2011
- Background current elimination in thin layer ion-selective membrane coulometryElectrochemistry Communications, 2010
- Thin Layer Coulometry with Ionophore Based Ion-Selective MembranesAnalytical Chemistry, 2010
- Rapid and coulometric electrolysis for ion transfer at the aqueous|organic solution interfaceJournal of Electroanalytical Chemistry, 2005
- Chronocoulometric flow-injection analysis with solvent polymeric membrane ion sensorsAnalytica Chimica Acta, 2002
- An ISFET-based microlitre titrator: integration of a chemical sensor—actuator systemSensors and Actuators, 1985