Tissue-Like 3D Standard and Protocols for Microscope Quality Management
- 4 March 2023
- journal article
- research article
- Published by Oxford University Press (OUP) in Microscopy and Microanalysis
- Vol. 29 (2) , 616-634
Abstract
This article outlines a global study conducted by the Association of Biomedical Resource Facilities (ABRF) Light Microscopy Research Group (LMRG). The results present a novel 3D tissue-like biologically relevant standard sample that is affordable and straightforward to prepare. Detailed sample preparation, instrument-specific image acquisition protocols and image analysis methods are presented and made available to the community. The standard consists of sub-resolution and large well characterized relative intensity fluorescence microspheres embedded in a 120 µm thick 3D gel with a refractive index of 1.365. The standard allows the evaluation of several properties as a function of depth. These include the following: 1) microscope resolution with automated analysis of the point-spread function (PSF), 2) automated signal-to-noise ratio analysis, 3) calibration and correction of fluorescence intensity loss, and 4) quantitative relative intensity. Results demonstrate expected refractive index mismatch dependent losses in intensity and resolution with depth, but the relative intensities of different objects at similar depths are maintained. This is a robust standard showing reproducible results across laboratories, microscope manufacturers and objective lens types (e.g., magnification, immersion medium). Thus, these tools will be valuable for the global community to benchmark fluorescence microscopes and will contribute to improved scientific rigor and reproducibility.Keywords
All Related Versions
This publication has 43 references indexed in Scilit:
- ConfocalCheck - A Software Tool for the Automated Monitoring of Confocal Microscope PerformancePLOS ONE, 2013
- Method of calibration of a fluorescence microscope for quantitative studiesJournal of Microscopy, 2011
- Quality Assurance Testing for Modern Optical Imaging SystemsMicroscopy and Microanalysis, 2011
- Accuracy and precision in quantitative fluorescence microscopyJournal of Cell Biology, 2009
- Calibrating excitation light fluxes for quantitative light microscopy in cell biologyNature Protocols, 2008
- Evaluating performance in three‐dimensional fluorescence microscopyJournal of Microscopy, 2007
- Quality assessment of confocal microscopy slide based systems: PerformanceCytometry Part A, 2006
- Effects of specimen refractive index on confocal imagingJournal of Microscopy, 1997
- Volume measurements in three‐dimensional microscopyScanning, 1994
- Aberrations in confocal fluorescence microscopy induced by mismatches in refractive indexJournal of Microscopy, 1993