Delay Line Readouts for High Purity Germanium Medical Imaging Cameras
Authors:
- 1 January 1974
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Nuclear Science
- Vol. 21 (1) , 652-657
Abstract
High purity germanium offers excellent potential for use in nuclear medical imaging cameras. A position and energy readout technique using two inexpensive delay lines has been developed for these cameras. Results obtained with a 1-cm2 , 4 mm deep, 5×5 strip high purity germanium detector are 2.1 mm full width spatial resolution, a measured single strip resolution of 0.65 mm full width half maximum (FWHM), a 25 element uncollimated energy resolution of 2.95 keV FWHM, and 2.65 keV FWHM for a single central element at 140 keV.Keywords
This publication has 4 references indexed in Scilit:
- Aspects of Imaging and Counting in Nuclear Medicine Using Scintillation and Semiconductor DetectorsIEEE Transactions on Nuclear Science, 1972
- An Experimental Comparison of Scintillation and Semiconductor Detectors for Isotope Imaging and CountingIEEE Transactions on Nuclear Science, 1972
- A Multi-Wire Proportional Chamber for Nuclear Medicine ApplicationsIEEE Transactions on Nuclear Science, 1972
- Clinical tests on a prototype semiconductor gamma-cameraBritish Journal of Radiology, 1971