Six years’ experience of using the BacT/ALERT system to screen all platelet concentrates, and additional testing of outdated platelet concentrates to estimate the frequency of false‐negative results
- 17 February 2005
- journal article
- Published by Wiley in Vox Sanguinis
- Vol. 88 (2) , 93-97
Abstract
Background and Objectives Approximately 1 in every 2000 units of platelets is contaminated with bacteria. The BacT/ALERT automated blood culture system can be used to screen platelet concentrates (PCs) for bacterial contamination.Materials and Methods Data were collected from May 1998 until May 2004. The number of PCs tested during this period was 36 896, most of which were produced from pools of four buffy‐coats. On the day following blood collection or platelet apheresis, a 5–10 ml sample of the PC was aseptically transferred to a BacT/ALERT culture bottle for detection of aerobic bacteria. The sample was monitored for bacterial growth during the entire storage period of the PC (6·5 days). When a positive signal was generated, the culture bottle, the PC and the erythrocyte concentrates were tested for bacterial growth. In order to determine the frequency of false‐negative BacT/ALERT signals, 1061 outdated PCs were tested during the period from May 2002 to May 2004.Results Eighty‐eight positive signals were detected by the BacT/ALERT system, of which 12 were interpreted as truly positive. Fourteen signals were interpreted as truly false positive. Thirty‐three signals were interpreted to be probably false positive. Two of 1061 outdated units tested positive, and Bacillus spp. and Staphylococcus epidermidis, respectively, were isolated from these PCs.Conclusions Between 0·03% and 0·12% of the PCs were contaminated with bacteria. BacT/ALERT is an efficient tool for monitoring PCs for bacterial contamination; however, it is important to realize that false‐negative results may occur.Keywords
This publication has 16 references indexed in Scilit:
- Positive BacT/ALERT signal from a platelet concentrate 23 h before the donor was admitted to hospital with Streptococcus pneumoniae septicaemiaVox Sanguinis, 2004
- Improving the bacteriological safety of platelet transfusionsTransfusion Medicine Reviews, 2004
- Monitoring of apheresis platelet bacterial contamination with an automated liquid culture system: a university experienceTransfusion, 2003
- Hemovigilance network in France: organization and analysis of immediate transfusion incident reports from 1994 to 1998Transfusion, 2002
- Transfusion‐transmitted bacterial infectionin the United States, 1998 through 2000Transfusion, 2001
- Evaluation of an automated culture systemfor detecting bacterial contamination of platelets: an analysis with 15 contaminating organismsTransfusion, 2001
- Current status of microbial contamination of blood components: summary of a conferenceTransfusion, 1997
- Sterility Testing of Blood Products in 1994/1995 by Three Cooperating Blood Banks in The NetherlandsVox Sanguinis, 1997
- Septic Reactions to Platelet TransfusionsJAMA, 1991
- Septic reactions to platelet transfusions. A persistent problem1991