Randomised short-term trial of high-span versus low-span APAP for treating sleep apnoea
- 12 June 2015
- journal article
- research article
- Published by Springer Nature in Sleep and Breathing
- Vol. 20 (1) , 183-190
Abstract
Auto-titrating continuous positive airway pressure (APAP) devices were developed to improve treatment efficacy and compliance in patients with obstructive sleep apnoea syndrome (OSAS). Since there are insufficient data on the optimal pressure range setting, we aimed to compare the adherence, efficacy and tolerability of treatment with high-span versus low-span APAP. Seventy-six newly diagnosed OSAS patients fulfilling the treatment criteria were randomised to receive high-span (HS, range 4–15cmH2O, n = 38) or low-span (LS, range 8–12cmH2O, n = 38) APAP. Patients were assessed at 1 and 3 months. Median Epworth sleepiness scale (ESS) was 13 (IQR, 6–16) and median apnoea-hypopnoea index (AHI) was 35.9 (IQR, 27.6–56.3). There were no significant differences in baseline demographic and clinical characteristics between groups. Overall, no significant differences were found at the first month assessment. After 3 months of therapy, we found again no differences in residual AHI or ESS. However, the group HS proved less adherent than group LS, respectively, with median 87 % (IQR, 60.5–97.5) versus 94 % (IQR, 80.0–98.3) of the nights using ≥4 h (P = 0.014) and mean (±SD) usage 5.7 ± 1.6 versus 6.4 ± 1.2 h/night (P = 0.049). The group HS reported more frequently nasal congestion, excessive oronasal dryness and nocturnal awakenings of at least moderate intensity, the latter with statistical significance (P = 0.005). Both pressure ranges appear to be equally effective to correct AHI and to improve symptoms. Though, patients with high-span APAP were less compliant to treatment, raising issues about the tolerability of wide pressure range settings of these devices.Keywords
This publication has 29 references indexed in Scilit:
- Auto-titrating versus fixed continuous positive airway pressure for the treatment of obstructive sleep apnea: a systematic review with meta-analysesSystematic Reviews, 2012
- Air Leak is Associated with Poor Adherence to Autopap TherapySLEEP, 2011
- Pressure modification for improving usage of continuous positive airway pressure machines in adults with obstructive sleep apnoeaCochrane Database of Systematic Reviews, 2009
- Cardiovascular outcomes of CPAP therapy in obstructive sleep apnea syndromeAmerican Journal of Physiology-Regulatory, Integrative and Comparative Physiology, 2007
- Bench Evaluation of Flow Limitation Detection by Automated Continuous Positive Airway Pressure DeviceChest, 2006
- Bench testing of auto-adjusting positive airway pressure devicesEuropean Respiratory Journal, 2004
- Randomised placebo controlled trial of daytime function after continuous positive airway pressure (CPAP) therapy for the sleep apnoea/hypopnoea syndromeThorax, 1998
- Automatic control of airway pressure for treatment of obstructive sleep apneaIEEE Transactions on Biomedical Engineering, 1995
- Obstructive sleep apnoea.Thorax, 1993
- REVERSAL OF OBSTRUCTIVE SLEEP APNOEA BY CONTINUOUS POSITIVE AIRWAY PRESSURE APPLIED THROUGH THE NARESThe Lancet, 1981