Rapid Quantitative Characterization of Ultrasonic Fields Using Fizeau Interferometry

Authors:

Abstract
The accurate characterization of air-coupled high-power ultrasonic fields is essential for advancing the development and control of ultrasonic transducer arrays. Conventional characterization methods often fail to provide both contactless visualization and a quantitative sound pressure distribution without the need for time-consuming point-by-point scanning. Acoustic pressure variations modulate the refractive index of the medium, e.g., ambient air, and thereby change the optical path length, which can be measured interferometrically. In this paper, Fizeau interferometry is systematically compared with Schlieren imaging and optical microphone measurements. The results demonstrate strong agreement with the reference methods and confirm the reliability of the interferometric approach. Overall, this method enables contactless two-dimensional field visualization with a spatial resolution of 23 µm in combination with quantitative sound pressure data within a single acquisition (40 ms). Consequently, it provides a rapid tool for characterizing high-power ultrasonic fields in applications such as acoustic levitation or sono-photonic systems.