Transient rise in His‐lead threshold due to acute myocardial infarction
- 19 January 2019
- journal article
- case report
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 42 (6) , 754-757
Abstract
An 85‐year‐old male was admitted to our center with a non‐ST elevation myocardial infarction. The patient had a dual‐chamber pacemaker in situ with an atrial and His lead. A transient increase in His threshold and loss of nonselective capture occurred at the presentation of right coronary artery infarction, peaking during rotational atherectomy therapy causing loss of capture and complete atrioventricular block. A follow‐up interrogation, 2 weeks postrevascularization, showed a return to a normal nonselective capture morphology and threshold measurements. Physicians should be aware of this complication in patients with His leads, particularly those with a history of coronary artery disease.Keywords
This publication has 13 references indexed in Scilit:
- Permanent His-bundle pacing: Long-term lead performance and clinical outcomesHeart Rhythm, 2018
- His Bundle Pacing: A New Frontier in the Treatment of Heart FailureArrhythmia and Electrophysiology Review, 2018
- Heart failure due to right ventricular apical pacing: the importance of flow patternsEP Europace, 2016
- Heart Block and Temporary Pacing During Rotational AtherectomyCanadian Journal of Cardiology, 2015
- Comparison of Frequency of Complications With On-Label Versus Off-Label Use of Rotational AtherectomyThe American Journal of Cardiology, 2012
- Anatomical Aspects of the Arterial Blood Supply to the Sinoatrial and Atrioventricular Nodes of the Human HeartJournal of International Medical Research, 2008
- Dual-Chamber Pacing or Ventricular Backup Pacing in Patients With an Implantable DefibrillatorJAMA, 2002
- Permanent, Direct His-Bundle PacingCirculation, 2000
- Normal Blood Supply to the Human His Bundle and Proximal Bundle BranchesCirculation, 1973
- Pervenous Pacing of the Specialized Conducting System in ManCirculation, 1970