Automatic contouring QA method using a deep learning–based autocontouring system
Authors:
Open Access
- 17 May 2022
- journal article
- research article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 23 (8) , e13647
Abstract
Purpose: To determine the most accurate similarity metric when using an independent system to verify automatically generated contours.Methods: A reference autocontouring system (primary system to create clinical contours) and a verification autocontouring system (secondary system to test the primary contours) were used to generate a pair of 6 female pelvic structures (UteroCervix [uterus + cervix], CTVn [nodal clinical target volume (CTV)], PAN [para‐aortic lymph nodes], bladder, rectum, and kidneys) on 49 CT scans from our institution and 38 from other institutions. Additionally, clinically acceptable and unacceptable contours were manually generated using the 49 internal CT scans. Eleven similarity metrics (volumetric Dice similarity coefficient (DSC), Hausdorff distance, 95% Hausdorff distance, mean surface distance, and surface DSC with tolerances from 1 to 10 mm) were calculated between the reference and the verification autocontours, and between the manually generated and the verification autocontours. A support vector machine (SVM) was used to determine the threshold that separates clinically acceptable and unacceptable contours for each structure. The 11 metrics were investigated individually and in certain combinations. Linear, radial basis function, sigmoid, and polynomial kernels were tested using the combinations of metrics as inputs for the SVM.Results: The highest contouring error detection accuracies were 0.91 for the UteroCervix, 0.90 for the CTVn, 0.89 for the PAN, 0.92 for the bladder, 0.95 for the rectum, and 0.97 for the kidneys and were achieved using surface DSCs with a thickness of 1, 2, or 3 mm. The linear kernel was the most accurate and consistent when a combination of metrics was used as an input for the SVM. However, the best model accuracy from the combinations of metrics was not better than the best model accuracy from a surface DSC as an input.Conclusions: We distinguished clinically acceptable contours from clinically unacceptable contours with an accuracy higher than 0.9 for the targets and critical structures in patients with cervical cancer; the most accurate similarity metric was surface DSC with a thickness of 1, 2, or 3 mm.Keywords
This publication has 18 references indexed in Scilit:
- Automated OAR Anomaly and Error Detection Tool in Radiation TherapyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2017
- V-Net: Fully Convolutional Neural Networks for Volumetric Medical Image Segmentation2016
- Fully Convolutional Networks for Semantic SegmentationIEEE Transactions on Pattern Analysis and Machine Intelligence, 2016
- Automated contouring error detection based on supervised geometric attribute distribution models for radiation therapy: A general strategyMedical Physics, 2015
- Groupwise Conditional Random Forests for Automatic Shape Classification and Contour Quality Assessment in Radiotherapy PlanningIEEE Transactions on Medical Imaging, 2013
- Comparison and Evaluation of Methods for Liver Segmentation From CT DatasetsIEEE Transactions on Medical Imaging, 2009
- Support vector machinesIEEE Intelligent Systems and their Applications, 1998
- Support-vector networksMachine Learning, 1995
- A training algorithm for optimal margin classifiers1992
- Grundzüge der MengenlehreMonatshefte für Mathematik, 1915