Differential pulse oximetry readings between ethnic groups and delayed transfer to intensive care units
Open Access
- 6 September 2022
- journal article
- research article
- Published by Oxford University Press (OUP) in QJM: An International Journal of Medicine
- Vol. 116 (1) , 63-67
Abstract
Pulse oximeters are widely used to monitor blood oxygen saturations, although concerns exist that they are less accurate in individuals with pigmented skin. This study aimed to determine if patients with pigmented skin were more severely unwell at the period of transfer to intensive care units (ICUs) than individuals with White skin. Using data from a large teaching hospital, measures of clinical severity at the time of transfer of patients with COVID-19 infection to ICUs were assessed, and how this varied by ethnic group. Data were available on 748 adults. Median pulse oximetry demonstrated similar oxygen saturations at the time of transfer to ICUs (Kruskal–Wallis test, P = 0.51), although median oxygen saturation measurements from arterial blood gases at this time demonstrated lower oxygen saturations in patients classified as Indian/Pakistani ethnicity (91.6%) and Black/Mixed ethnicity (93.0%), compared to those classified as a White ethnicity (94.4%, Kruskal–Wallis test, P = 0.005). There were significant differences in mean respiratory rates in these patients (P < 0.0001), ranging from 26 breaths/min in individuals with White ethnicity to 30 breaths/min for those classified as Indian/Pakistani ethnicity and 31 for those who were classified as Black/Mixed ethnicity. These data are consistent with the hypothesis that differential measurement error for pulse oximeter readings negatively impact on the escalation of clinical care in individuals from other than White ethnic groups. This has implications for healthcare in Africa and South-East Asia and may contribute to differences in health outcomes across ethnic groups globally.Keywords
Funding Information
- Nottingham University Hospitals NHS Trust
- University of Nottingham
This publication has 15 references indexed in Scilit:
- Racial Bias in Pulse Oximetry MeasurementNew England Journal of Medicine, 2020
- Changes in Racial Disparities in Mortality After Cancer Surgery in the US, 2007-2016JAMA Network Open, 2020
- Race, Postoperative Complications, and Death in Apparently Healthy ChildrenPediatrics, 2020
- Racial Disparities In Surgical Mortality: The Gap Appears To Have NarrowedHealth Affairs, 2017
- Pulse oximetryCritical Care, 2015
- Racial and ethnic disparities in mortality from acute lung injury*Critical Care Medicine, 2009
- The widening black/white gap in asthma hospitalizations and mortalityJournal of Allergy and Clinical Immunology, 2006
- Effects of Skin Pigmentation on Pulse Oximeter Accuracy at Low SaturationAnesthesiology, 2005
- Reliability of Pulse Oximetry in Titrating Supplemental Oxygen Therapy in Ventilator-Dependent PatientsChest, 1990
- The Environment and Disease: Association or Causation?Proceedings of the Royal Society of Medicine, 1965