Plasma Sterilisation within Long and Narrow Bore Dielectric Tubes Contaminated with Stacked Bacterial Spores
Authors:
- 21 December 2007
- journal article
- research article
- Published by Wiley in Plasma Processes and Polymers
- Vol. 5 (1) , 14-25
Abstract
Catheters, which comprise small diameter (≤4 mm), long (typically more than a meter), thermosensitive polymer tubings, are generally used only once. This is because conventional low temperature sterilisation techniques are considered inadequate for used catheters. The plasma sterilisation method proposed in the current paper should allow for the achievement of re‐sterilisation of catheters according to accepted regulations. The plasma process described enables one to sterilise, in less than 10 min, the inner part of a long 4 mm i.d. Teflon tube contaminated initially with 106 Bacillus atrophaeus spores. This result was obtained by achieving an argon discharge at reduced pressure (750 mTorr) within the hollow (dielectric) tube itself. The discharge was sustained using a microwave field‐applicator called a stripline, fully enclosing the tube to be treated. This linear field‐applicator yields a uniform plasma all along the tube, hence the uniform biocide action. The biocide agents are the vacuum ultra‐violet (VUV) photons, which include oxygen and nitrogen atomic lines, the N2 Lyman‐Birge‐Hopfield (LBH) bands, the UV photons emitted by the NOβ and NOγ molecular systems resulting from the contamination, even though at a very low‐level, of the argon gas (high purity argon is used) by air. Scanning electron microscopy (SEM) revealed no apparent damage to the external structure of the spores and to polystyrene microspheres exposed to plasma during the time required for reaching sterility.To check for sterility in such narrow bore tubes without having to cut them into two pieces, a procedure was developed to introduce and afterwards collect the bacterial spores used as bio‐indicators. This diagnostic procedure allowed, at the same time, the imaging of the microorganisms relatively efficiently with SEM, showing the eventual stacking of bacterial spores, a possible source of sterilisation failure. magnified imageKeywords
This publication has 15 references indexed in Scilit:
- Long and uniform plasma columns generated by linear field-applicators based on stripline technologyPlasma Sources Science and Technology, 2007
- Sterility and Microbiological Assessment of Reused Single-Use Cardiac Electrophysiology CathetersInfection Control & Hospital Epidemiology, 2006
- Silcone Catheter Sterilization by Microwave Plasma; Argon and Nitrogen DischargeInternational Journal of Infrared and Millimeter Waves, 2005
- The respective roles of UV photons and oxygen atoms in plasma sterilization at reduced gas pressure: the case of N/sub 2/-O/sub 2/ mixturesIEEE Transactions on Plasma Science, 2002
- Plasma sterilization. Methods and mechanismsPure and Applied Chemistry, 2002
- Reclassification of bioindicator strains Bacillus subtilis DSM 675 and Bacillus subtilis DSM 2277 as Bacillus atrophaeus.International Journal of Systematic and Evolutionary Microbiology, 2001
- The Presterilization Microbial Load on Used Medical Devices and the Effectiveness of Hydrogen Peroxide Gas Plasma AgainstBacillus SubtilisSporesInfection Control & Hospital Epidemiology, 1999
- Comparison of Ion Plasma, Vaporized Hydrogen Peroxide, and 100% Ethylene Oxide Sterilizers to the 12/88 Ethylene Oxide Gas SterilizerInfection Control & Hospital Epidemiology, 1996
- Plasma sources using long linear microwave field applicators: main features, classification and modellingPlasma Sources Science and Technology, 1995
- Sterilization of various diameter dead‐ended tubesBiotechnology and Bioengineering, 1993